A M Scholtens1, L E Swart2, H J Verberne3, R P J Budde2, M G E H Lam4.
Abstract
PURPOSE: FDG PET/CT has been of increasing interest in the diagnostic workup of prosthetic heart valve endocarditis (PVE). Some reports advocate later imaging time points to improve the diagnostic accuracy for PVE. In this study, we compared standard and late FDG PET/CT images in patients with a clinical suspicion of PVE.Entities:
Keywords: FDG PET/CT; endocarditis; prosthetic heart valve
Mesh:
Substances:
Year: 2017 PMID: 28474192 PMCID: PMC6280957 DOI: 10.1007/s12350-017-0902-3
Source DB: PubMed Journal: J Nucl Cardiol ISSN: 1071-3581 Impact factor: 5.952
Patient characteristics
| Patient | Age | Valve | Days post implant | Implantation/complications | CRP | Leukocyte count | PVE | |
|---|---|---|---|---|---|---|---|---|
| Type | Location | |||||||
| 1 | 72 | Mechanical | MV | 4992 | Uncomplicated implantation | 43 | 10.1 | No |
| 2 | 46 | Mechanical | AV | 28 | With Bentall-graft. Postoperative haematoma requiring re-thoracotomy and re-suturing of the graft | 9 | 7.8 | No |
| 3 | 65 | Mechanical | AV | 645 | Uncomplicated implantation | 71 | 10.5 | No |
| 4 | 84 | Biological | AV | 21 | Uncomplicated implantation | 154 | 11.7 | No |
| 5 | 52 | Mechanical | AV | 922 | Postoperative pneumothorax. No complications of the valve | 184 | 12.7 | No |
| 6 | 83 | Biological | AV | 63 | Uncomplicated implantation. CABG in same session | 10 | 7.5 | No |
| 7 | 71 | Biological | AV | 150 | Uncomplicated implantation | 89 | 11.1 | No |
| 8 | 57 | Mechanical | AV | 2694 | Uncomplicated implantation | 35 | 7.8 | No |
| 9 | 49 | Biological | AV | 1769 | With Bentall-graft. Uncomplicated implantation | 35 | 11.5 | Yes |
| 10 | 75 | Biological | AV | 143 | Postoperative pneumothorax. No complications of the valve. CABG in same session. | 43 | 7.6 | Yes |
| 11 | 78 | Biological | AV | 101 | Uncomplicated implantation. CABG in same session | 2 | 4.1 | Yes |
| 12 | 23 | Biological | PV | 2501 | Uncomplicated implantation | 43 | 8.2 | Yes |
| 13 | 68 | Biological | AV | 662 | Uncomplicated implantation | 30 | 9.5 | Yes |
| 710 | 8 | 7.9 | Yes | |||||
Contingency tables of the visual analysis
| Standard | Late | |||
|---|---|---|---|---|
| PET/CT+ | PET/CT− | PET/CT+ | PET/CT− | |
| PVE+ | 5 | 1 | 6 | 0 |
| PVE− | 1 | 7 | 5 | 3 |
Figure 1Semi-quantitative measurements. Values for standard SUVmax (A), late SUVmax (B), changes in SUVmax from standard to late (ΔSUVmax, C) as well as standard and late Target-to-background ratios (TBR) and changes in TBR (D, E, and F respectively) with boxplot representations of median values, interquartile ranges, and total ranges. * Statistical outlier
Figure 2Examples of standard (top row) and late (bottom row) FDG PET/CT images in six patients. A True positive uptake surrounding a mechanical aortic PHV in both phases. B True negative images with hardly any uptake near the biological aortic PHV in either phase. C True negative standard images with minimal uptake surrounding the mechanical mitral PHV, clearly delineated in the late image. D True positive late phase with focal uptake near the strut of the biological aortic PHV near the right coronary that is not apparent on standard images. Note accumulation of uptake in mediastinal lymph nodes. E False positive late images with exaggerated uptake near all struts of a biological aortic PHV with normal uptake on standard images. F False positive uptake in both phases surrounding the prosthesis after semi-recent Bentall procedure complicated by haematoma. Note the unsuppressed physiological uptake in the myocardium