Hai-Jian Zhao1, Tian-Jiao Zhang1, Jie Zeng1, Cui-Hua Hu1, Rong Ma1, Chuan-Bao Zhang1.
Abstract
BACKGROUND: Hemoglobin A1c (HbA1c) measurement is of great value for the diagnosis and monitoring of diabetes. Many manufacturers have developed various experiments to determine the HbA1c concentration. However, the longitudinal use of these tests requires strict quality management. This study aimed to analyze the quality of HbA1c measurement systems in China using six sigma techniques to help improve their performances.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28469104 PMCID: PMC5421179 DOI: 10.4103/0366-6999.204932
Source DB: PubMed Journal: Chin Med J (Engl) ISSN: 0366-6999 Impact factor: 2.628
Principles, instruments, reagents, and calibrators used by all laboratories in this study
| Principles and instruments | Reagents | Calibrators | Number of laboratories |
|---|---|---|---|
| Automated cation exchange high-performance liquid chromatography | |||
| ARKRAY HA-8160 | ARKARY | ARKARY | 10 |
| ARKRAY HA-8180 | ARKARY | ARKARY | 10 |
| BIO-RAD D-10 | Bio-Rad | Bio-Rad | 26 |
| BIO-RAD Variant II | Bio-Rad | Bio-Rad | 17 |
| BIO-RAD Variant II Turbo | Bio-Rad | Bio-Rad | 17 |
| TOSOH G7 | Tosoh | Tosoh | 4 |
| TOSOH G8 | Tosoh | Tosoh | 27 |
| Hui-zhong MQ-2000 PT | Hui-zhong | Hui-zhong | 4 |
| Automated affinity chromatography HPLC | |||
| PRIMUS HPLC | Primus | Bio-Rad | 1 |
| PRIMUS HPLC | Primus | PRIMUS | 5 |
| Enzymatic method | |||
| Beckman AU | First Chemistry | First Chemistry | 2 |
| Mindray BS | Mindray | Mindray | 1 |
| Immunoturbidimetry | |||
| Abbott architect | Simes Sikma | Simes Sikma | 1 |
| Beckman Synchron | Medical System | Medical System | 1 |
| Roche cobas | Roche | Roche | 4 |
| Roche others | Roche | Roche | 1 |
| Dade Behring Dimension | Siemens | Siemens | 1 |
| Hitachi | Medical System | Medical System | 1 |
| Hitachi | Randox | Randox | 1 |
| Hitachi | Zhe-jiang Dongou | Zhe-jiang Dongou | 1 |
HPLC: High-performance liquid chromatography.
Laboratories meeting bias criteria, imprecision criteria, and both criteria in this study
| Principle and instrument groups | Number of laboratories | Laboratories meeting different criteria, | ||
|---|---|---|---|---|
| Biasa | CVa | Both biasa and CVa | ||
| HPLC | 121 | 97 (80.2) | 76 (62.8) | 64 (52.9) |
| BIO-RAD | 60 | 53 (88.3) | 36 (60.0) | 32 (53.3) |
| TOSOH | 31 | 26 (83.9) | 22 (71.0) | 19 (61.3) |
| ARKRAY | 20 | 11 (55.0) | 13 (65.0) | 8 (40.0) |
| PRIMUS | 6 | 3 (50.0) | 3 (50.0) | 3 (50.0) |
| Hui-zhong | 4 | 4 (100.0) | 2 (50.0) | 2 (50.0) |
| Enzymatic method | 3 | 2 (66.7) | 2 (66.7) | 2 (66.7) |
| Immunoturbidimetry | 11 | 5 (45.5) | 6 (54.6) | 3 (27.3) |
| All | 135 | 104 (77.0) | 84 (62.2) | 69 (51.1) |
HPLC: High-performance liquid chromatography; Biasa: Allowable bias; CVa: Allowable coefficient of variation.
Constituent ratios of sigma values for different groups in this study
| Principle and instrument groups | Number of laboratories | Number of laboratories (%) | ||||
|---|---|---|---|---|---|---|
| σ<3 | 3≤σ<4 | 4≤σ<5 | 5≤σ<6 | σ≥6 | ||
| HPLC | 121 | 80 (66.1) | 16 (13.2) | 10 (8.3) | 7 (5.8) | 8 (6.6) |
| BIO-RAD | 60 | 43 (71.7) | 8 (13.3) | 5 (8.3) | 4 (6.7) | (0.0) |
| TOSOH | 31 | 15 (48.4) | 4 (12.9) | 5 (16.1) | 2 (6.5) | 5 (16.1) |
| ARKRAY | 20 | 14 (70.0) | 2 (10.0) | (0.0) | 1 (5.0) | 3 (15.0) |
| PRIMUS | 6 | 5 (83.3) | 1 (16.7) | (0.0) | (0.0) | (0.0) |
| Hui-zhong | 4 | 3 (75.0) | 1 (25.0) | (0.0) | (0.0) | (0.0) |
| Enzymatic method | 3 | 1 (33.3) | 1 (33.3) | (0.0) | (0.0) | 1 (33.3) |
| Immunoturbidimetry | 11 | 7 (69.6) | 1 (9.1) | 1 (9.1) | (0.0) | 2 (18.2) |
| All | 135 | 88 (65.2) | 18 (13.3) | 11 (8.2) | 7 (5.2) | 11 (8.2) |
HPLC: High-performance liquid chromatography.
Constituent ratios of QGI for laboratories with σ<3 in this study
| Principle and instrument groups | Number of laboratories | Number of laboratories (%) | ||
|---|---|---|---|---|
| QGI <0.8 | 0.8≤ QGI ≤1.2 | QGI >1.2 | ||
| HPLC | 80 | 47 (58.8) | 13 (16.3) | 20 (25.0) |
| BIO-RAD | 43 | 33 (76.8) | 3 (7.0) | 7 (16.3) |
| TOSOH | 15 | 9 (60.0) | 1 (6.7) | 5 (33.3) |
| ARKRAY | 14 | 3 (21.4) | 5 (35.7) | 6 (42.9) |
| PRIMUS | 5 | 1 (20.0) | 2 (40.0) | 2 (40.0) |
| Hui-zhong | 3 | 1 (33.3) | 2 (66.7) | (0.0) |
| Enzymatic method | 1 | 1 (100.0) | (0.0) | (0.0) |
| Immunoturbidimetry | 7 | 4 (57.1) | (0.0) | 3 (42.9) |
| All | 88 | 52 (59.1) | 13 (14.8) | 23 (26.1) |
HPLC: High-performance liquid chromatography; QGI: Quality goal index.