Real Structural Valve Deterioration of the Mitroflow Aortic Prosthesis: Competing Risk Analysis.

Rocío Díaz1, Daniel Hernández-Vaquero2, Jacobo Silva3, Isaac Pascual2, Jesús M de la Hera2, Víctor León2, María Martín2, Vicente Barriales2, Santiago Colunga2, Raquel Del Valle2, César Morís3.   

Abstract

INTRODUCTION AND
OBJECTIVES: The Mitroflow aortic prosthesis is a bovine pericardial bioprosthesis specially designed to increase the valve area in relation to its size. There is controversy regarding the pattern of structural valve deterioration (SVD). Our aim was to determine the cumulative incidence of SVD, risk factors influencing its occurrence, and its impact on mortality.
METHODS: A total of 1028 patients were clinically and echocardiographically followed up. Because the study population was elderly and had heart disease, we used a competing risk analysis.
RESULTS: The percentage of patients with SVD at 5 years was 4.22% (95%CI, 2.96-5.81) and was 15.77% at 8 years (95%CI, 12.46-19.43). The incidence was higher for small valves (19mm and 21mm) reaching 6.43% at 5 years (95%CI, 4.48-8.84) and 20.06% at 8 years (95%CI, 15.53-25.01). Severe patient-prosthesis mismatch (PPM) influenced the incidence of SVD (sHR, 3.53; 95%CI, 2.20-5.66; P < .001) but moderate PPM had no impact. The most powerful predictor of mortality was the presence of SVD (HR, 4.59; 95%CI, 2.91-7.22; P < .001).
CONCLUSIONS: This study used a definition based on the increase in the transprosthetic gradient and found a higher incidence of SVD of the Mitroflow prosthesis than that reported by other series, especially for sizes 19mm and 21mm and in patients with severe PPM. The incidence of SVD increased exponentially from the fifth year after implantation and its occurrence led to a 4.5-fold increase in the risk of death.
Copyright © 2017 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.

Entities:  

Keywords:  Aortic valve replacement; Bioprosthesis; Bioprótesis; Degeneración estructural valvular; Recambio valvular aórtico; Structural valve deterioration

Mesh:

Year:  2017        PMID: 28465144     DOI: 10.1016/j.rec.2017.02.041

Source DB:  PubMed          Journal:  Rev Esp Cardiol (Engl Ed)        ISSN: 1885-5857


  1 in total

1.  Long term follow up of percutaneous treatment for degenerated Mitroflow prosthesis with self-expanding transcatheter aortic valve implantation.

Authors:  Isaac Pascual; Marcel Almendárez; Rut Álvarez Velasco; Antonio Adeba; Daniel Hernández-Vaquero; Rebeca Lorca; Rocío Díaz; Alberto Alperi; Héctor Cubero-Gallego; Jose Rozado; César Morís; Pablo Avanzas
Journal:  Ann Transl Med       Date:  2020-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.