| Literature DB >> 28464517 |
Jennifer B Smilowitz1,2, David Dunkerley2, Patrick M Hill1, Poonam Yadav1, Mark W Geurts1.
Abstract
The dosimetric stability of six TomoTherapy units was analyzed to investigate changes in performance over time and with system upgrades. Energy and output were tracked using monitor chamber signal, onboard megavoltage computed tomography (MVCT) detector profile, and external ion chamber measurements. The systems (and monitoring periods) include three Hi-Art (67, 61, and 65 mos.), two TomoHDA (31 and 26 mos.), and one Radixact unit (11 mos.), representing approximately 10 years of clinical use. The four newest systems use the Dose Control Stability (DCS) system and Fixed Target Linear Accelerator (linac) (FTL). The output stability is reported as deviation from reference monitor chamber signal for all systems and/or from an external chamber signal. The energy stability was monitored using relative (center versus off-axis) MVCT detector signal (beam profile) and/or the ratio of chamber measurements at 2 depths. The clinical TomoHDA data were used to benchmark the Radixact stability, which has the same FTL but runs at a higher dose rate. The output based on monitor chamber data of all systems is very stable. The standard deviation of daily output on the non-DCS systems was 0.94-1.52%. As expected, the DCS systems had improved standard deviation: 0.004-0.06%. The beam energy was also very stable for all units. The standard deviation in profile flatness was 0.23-0.62% for rotating target systems and 0.04-0.09% for FTL. Ion chamber output and PDD ratios supported these results. The output stability on the Radixact system during extended treatment delivery (20, 30, and 40 min) was comparable to a clinical TomoHDA system. For each system, results are consistent between different measurement tools and techniques, proving not only the dosimetric stability, but also these quality parameters can be confirmed with various metrics. The replacement history over extended time periods of the major dosimetric components of the different delivery systems (target, linac, and magnetron) is also reported.Entities:
Keywords: Radixact; dosimetric stability; tomotherapy
Mesh:
Year: 2017 PMID: 28464517 PMCID: PMC5689853 DOI: 10.1002/acm2.12085
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Systems studied and monitoring periods. All systems, except the Radixact research unit are clinical treatment units. The Hi‐Art SN 103 is listed twice because DCS was added and the data for this machine is considered with and without DCS
| TomoTherapy unit | Monitor chamber & exit detector (mos.) | Ion chamber (mos.) | Service records (mos.) |
|---|---|---|---|
| Non‐DCS 04, Hi‐Art | 67 | NA | 42 |
| Non‐DCS 96, Hi‐Art | 61 | NA | 34 |
| Non‐DCS 103, Hi‐Art | 65 | NA | 21 |
| DCS 103, Hi‐Art | 37 | 32 | 33 |
| DCS 477, TomoHDA | 31 | 27 | 31 |
| DCS 488, TomoHDA | 26 | 26 | 26 |
| DCS RS, Research | 3 | 8 | 11 |
Figure 1Exit Detector signals from 175 J48 RotVar procedures on SN 04. The RMS difference in the profile with respect to baseline is used a surrogate of beam energy. The characteristic shape of the flattening filter‐free beam is obscured in the center detector data due to the deliberate offset of the CT detector focal point from the beam source point leading to a lack of buildup in the center channels.
Major dosimetric component replacement history. The service record history for the Hi‐Art systems does not cover the complete duration of the output and energy study due to limited availability of older records
| Unit [time period (mos.)] | Magnetron | Linac | Target |
|---|---|---|---|
| Hi‐Art SN 04 (42) | 10 | 4 | 5 |
| Hi‐Art SN 96 (34) | 11 | 1 | 4 |
| Hi‐Art 103(21) | 4 | 0 | 3 |
| Hi‐Art 103 (33), new target & linac design | 6 | 0 | 1 |
| TomoHDA 477 (31) | 0 | 0 | 0 |
| TomoHDA 488 (26) | 1 | 0 | 0 |
| Radixact (11) | 0 | 0 | 0 |
The service record history does not cover the complete duration of this output and energy study due to limited availability of records for the Hi‐Art systems.
Output stability. The standard deviation of output and (% within 2σ of mean) for all machines for which the data was available. An A1SL was used as an external ion chamber to measure output in solid water phantom. The monitor chamber signal is from the primary internal chamber (dose1)
| Non‐DCS 04 | Non‐DCS 96 | Non‐DCS 103 | DCS 103 | DCS 477 | DCS 488 | DCS RS | |
|---|---|---|---|---|---|---|---|
| A1SL | N/A | N/A | 0.43 (93.3%) | 0.004 (94.1%) | 0.56 (100%) | 0.33 (92.9%) | 0.34 (100%) |
| Monitor chamber | 1.52 (97.3%) | 1.30 (96.8%) | 0.94 (96.8%) | 0.004 (93.8%) | 0.048 (97.4%) | 0.061 (98.7%) | 0.035 (100%) |
Figure 2Relative output for non‐DCS systems measured by the monitor chamber. Data for days on which the machine was being serviced or otherwise out of the daily ±3% tolerance (and therefore not acceptable for treatment) are not included.
Figure 3Relative output for DCS systems measured by the monitor chamber. Data for days on which the machine was being serviced or otherwise out of the daily ±3% tolerance (and therefore not acceptable for treatment) are not included.
Energy stability. The mean and (number of data points) of exit detector RMS flatness and PDD 20/10 ratio for all machines
| Rotating 04 | Rotating 96 | Rotating 103 | Fixed 103 | Fixed 477 | Fixed 488 | Fixed RS | |
|---|---|---|---|---|---|---|---|
| RMS flatness | 0.62% (175) | 0.23% (61) | 0.59% (236) | 0.09% (54) | 0.07% (662) | 0.09% (566) | 0.04% (20) |
| PDD 20/10 ratio | N/A | N/A | 0.18% (885) | N/A | 0.49% (12) | 0.72% (13) | 0.04% (67) |
Figure 4Comparisons of normalized exit detector ratios. Normalized exit detector ratios are illustrated as a surrogate for beam energy stability over time. The rotating target system (SN04, left) shows greater difference over all datasets compared to a fixed target (FTL) system (SN103, right) which shows less variation in profiles. Both datasets are displayed are on equal scales.
Extended duration delivery stability. The coefficient of variation (CoV) is reported for energy (ratio of 10 and 20 cm depth readings) and output (10 cm) during 20, 30 and 40 min. extended deliveries on Radixact and TomoHDA
| Plan duration [min] | CoV energy (%) | CoV 10 cm (%) | |
|---|---|---|---|
| Radixact | 20 | 0.062 | 0.079 |
| 30 | 0.071 | 0.087 | |
| 40 | 0.099 | 0.109 | |
| TomoHDA | 20 | 0.067 | 0.075 |
| 30 | 0.059 | 0.077 | |
| 40 | 0.071 | 0.094 |
The coefficient of variation (CoV) is reported for energy (ratio of 10 and 20 cm depth readings) and output (10 cm) during 20, 30, and 40 min. Extended deliveries on Radixact and TomoHDA.