Accuracy of MDCT and CBCT in three-dimensional evaluation of the oropharynx morphology.

Hui Chen1,2, Maureen van Eijnatten3, Ghizlane Aarab2, Tim Forouzanfar3, Jan de Lange4, Paul van der Stelt1, Frank Lobbezoo2, Jan Wolff3.   

Abstract

Objective: To assess the accuracy of five different computed tomography (CT) scanners for the evaluation of the oropharynx morphology.
Methods: An existing cone-beam computed tomography (CBCT) data set was used to fabricate an anthropomorphic phantom of the upper airway volume that extended from the uvula to the epiglottis (oropharynx) with known dimensions (gold standard). This phantom was scanned using two multi-detector row computed tomography (MDCT) scanners (GE Discovery CT750 HD, Siemens Somatom Sensation) and three CBCT scanners (NewTom 5G, 3D Accuitomo 170, Vatech PaX Zenith 3D). All CT images were segmented by two observers and converted into standard tessellation language (STL) models. The volume and the cross-sectional area of the oropharynx were measured on the acquired STL models. Finally, all STL models were registered and compared with the gold standard.
Results: The intra- and inter-observer reliability of the oropharynx segmentation was fair to excellent. The most accurate volume measurements were acquired using the Siemens MDCT (98.4%; 14.3 cm3) and Vatech CBCT (98.9%; 14.4 cm3) scanners. The GE MDCT, NewTom 5G CBCT, and Accuitomo CBCT scanners resulted in smaller volumes, viz., 92.1% (13.4 cm3), 91.5% (13.3 cm3), and 94.6% (13.8 cm3), respectively. The most accurate cross-sectional area measurements were acquired using the Siemens MDCT (94.6%; 282.4 mm2), Accuitomo CBCT (95.1%; 283.8 mm2), and Vatech CBCT (95.3%; 284.5 mm2) scanners. The GE MDCT and NewTom 5G CBCT scanners resulted in smaller areas, viz., 89.3% (266.5 mm2) and 89.8% (268.0 mm2), respectively. Limitations: Images of the phantom were acquired using the vendor-supplied default airway scanning protocol for each scanner.
Conclusion: Significant differences were observed in the volume and cross-sectional area measurements of the oropharynx acquired using different MDCT and CBCT scanners. The Siemens MDCT and the Vatech CBCT scanners were more accurate than the GE MDCT, NewTom 5G, and Accuitomo CBCT scanners. In clinical settings, CBCT scanners offer an alternative to MDCT scanners in the assessment of the oropharynx morphology.
© The Author 2017. Published by Oxford University Press on behalf of the European Orthodontic Society. All rights reserved. For permissions, please email: journals.permissions@oup.com

Entities:  

Mesh:

Year:  2018        PMID: 28453722     DOI: 10.1093/ejo/cjx030

Source DB:  PubMed          Journal:  Eur J Orthod        ISSN: 0141-5387            Impact factor:   3.075


  3 in total

1.  Geometric Validation of Continuous, Finely Sampled 3-D Reconstructions From aOCT and CT in Upper Airway Models.

Authors:  Hillel B Price; Julia S Kimbell; Ruofei Bu; Amy L Oldenburg
Journal:  IEEE Trans Med Imaging       Date:  2018-10-17       Impact factor: 10.048

2.  A new segmentation algorithm for measuring CBCT images of nasal airway: a pilot study.

Authors:  Chen Zhang; Robin Bruggink; Frank Baan; Ewald Bronkhorst; Thomas Maal; Hong He; Edwin M Ongkosuwito
Journal:  PeerJ       Date:  2019-01-28       Impact factor: 2.984

3.  Interobserver and intraobserver variability in the radiological assessment of sialolithiasis using cone beam computed tomography.

Authors:  L-J Beumer; E-H van der Meij; J-I Kamstra; J-G de Visscher
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2022-01-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.