Optimizing Contrast Media Injection Protocols in Computed Tomography Angiography at Different Tube Voltages: Evaluation in a Circulation Phantom.

Damiano Caruso1, Marwen Eid, U Joseph Schoepf, Domenico De Santis, Akos Varga-Szemes, Stefanie Mangold, Christian Canstein, Virginia W Lesslie, Stephen R Fuller, B Devon Ball, Andrea Laghi, Carlo N De Cecco.   

Abstract

OBJECTIVES: The aim of this study was to investigate the minimum iodine delivery rate (IDR) and contrast media (CM) volume required for diagnostic contrast enhancement of 350 HU (Hounsfield units) in the ascending aorta at different kV settings.
METHODS: Dynamic computed tomography acquisitions from 70 to 150 kV were performed in a circulation phantom. First, injections with IDR ranging from 0.1 to 2.0 g I/s were tested for each kV. In the second part, the IDR was held constant, whereas the CM volume was reduced from 50 to 10 mL. Diagnostic aortic peak enhancement for each kV was compared using the Kruskal-Wallis test. P < 0.05 was considered statistically significant.
RESULTS: The mean aortic peak enhancement for all diagnostic IDRs was 368.7 ± 11.1 HU. Diagnostic IDRs returned similar aortic peak enhancement values for all protocols (all P ≥ 0.18). For the second part of the study, a diagnostic enhancement was yielded by using a minimum of 30 mL of CM for 110 kV, 25 mL for 100 and 90 kV, and 15 mL for 80 and 70 kV.
CONCLUSION: Our study suggests that a differentiated approach reducing the CM volume for tube voltages of less than 120 kV and increasing the IDR for higher kV settings seems to be the most effective approach.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28448412     DOI: 10.1097/RCT.0000000000000613

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  5 in total

1.  Contrast media injection protocol optimization for dual-energy coronary CT angiography: results from a circulation phantom.

Authors:  Domenico De Santis; Damiano Caruso; U Joseph Schoepf; Marwen Eid; Moritz H Albrecht; Taylor M Duguay; Akos Varga-Szemes; Andrea Laghi; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2018-02-27       Impact factor: 5.315

2.  Impact of CT Injector Technology and Contrast Media Viscosity on Vascular Enhancement: Evaluation in a Circulation Phantom.

Authors:  Michael McDermott; Corey Kemper; William Barone; Gregor Jost; Jan Endrikat
Journal:  Br J Radiol       Date:  2020-02-20       Impact factor: 3.039

3.  Optimization of contrast medium volume for abdominal CT in oncologic patients: prospective comparison between fixed and lean body weight-adapted dosing protocols.

Authors:  Damiano Caruso; Elisa Rosati; Nicola Panvini; Marco Rengo; Davide Bellini; Giulia Moltoni; Benedetta Bracci; Elena Lucertini; Marta Zerunian; Michela Polici; Domenico De Santis; Elsa Iannicelli; Paolo Anibaldi; Iacopo Carbone; Andrea Laghi
Journal:  Insights Imaging       Date:  2021-03-20

4.  Impact of iodine concentration and scan parameters on image quality, contrast enhancement and radiation dose in thoracic CT.

Authors:  Marian S Solbak; Mette K Henning; Andrew England; Anne C Martinsen; Trond M Aaløkken; Safora Johansen
Journal:  Eur Radiol Exp       Date:  2020-09-11

5.  Design, Implementation, and Validation of a Pulsatile Heart Phantom Pump.

Authors:  Volkan Tuncay; Jan Zijlstra; Matthijs Oudkerk; Peter M A van Ooijen
Journal:  J Digit Imaging       Date:  2020-08-10       Impact factor: 4.056