| Literature DB >> 28441708 |
Alparslan Serarslan1, Nilgun Ozbek Okumus, Bilge Gursel, Deniz Meydan, Yalcin Dastan, Talat Aksu.
Abstract
Background: The current optimal radiotherapy (RT) planning technique for stomach cancer is controversial. The design of RT for stomach cancer is difficult and differs according to tumor localization. Dosimetric and clinical studies have been performed in patients with different tumor localizations. This may be the main source of inconsistencies in study results. For this reason, we attempted to find the optimal RT technique for patients with stomach cancer in similar locations.Entities:
Keywords: Conformal radiotherapy; dosimetry; field-in-field radiotherapy; intensity; modulated radiotherapy
Year: 2017 PMID: 28441708 PMCID: PMC5464493 DOI: 10.22034/APJCP.2017.18.3.741
Source DB: PubMed Journal: Asian Pac J Cancer Prev ISSN: 1513-7368
Dosimetric Parameters for Planning Target Volumes with Three Planning Techniques
| Mean±SD (minimum-maximum) | ||||||
|---|---|---|---|---|---|---|
| WB-CRT | FIF-IMRT | IMRT | WB-CRT vs. FIF-IMRT | WB-CRT vs. IMRT | FIF-MRT vs. IMRT | |
| Mean dose (cGy) | 4,640.95±27.74 | 4,606.80±22.84 | 4,657.05±27.67 | 0.0002 | 0.059 | 0.0001 |
| (4,601-4,709) | (4,563-4,648) | (4,611-4,729) | ||||
| Maximum dose (cGy) | 4,830.70±50.39 | 4,741.60±32.61 | 4,913.75±64.22 | 0.00008 | 0.001 | 0.00008 |
| (4,755-4,971) | (4,677-4,812) | (4,780-4,970) | ||||
| V95 (%) | 97.07±1.5 | 97.07±1.61 | 96.89±1.63 | 0.823 | 0.614 | 0.852 |
| (95.00-99.50) | (95.00-99.40) | (95.00-96.60) | ||||
| D2 (cGy) | 4,771.85±37.93 | 4,697.30±29.20 | 4,814.25±54.25 | 0.00008 | 0.006 | 0.00008 |
| (4,712-4,858) | (4,644-4,737) | (4,722-4,879) | ||||
| D98 (cGy) | 4,487.00±29.57 | 4,485.45±32.35 | 4,466.65±44.58 | 0.538 | 0.173 | 0.247 |
| (4,434-4,538) | (4,435-4,543) | (4,397-4,530) | ||||
| V > 107 (%) | 0.74±1.91 | 0.00 | 3.61±3.69 | 0.012 | 0.003 | 0.0002 |
| (0.00-7.82) | (0.00-0.00) | (0.00+11.35) | ||||
| V < 95 (%) | 2.87±1.58 | 2.88±1.60 | 3.10±1.63 | 0.823 | 0.575 | 0.808 |
| (0.50-5.00) | (0.60-5.61) | (0.40-5.00) | ||||
| Homogenity index | 6.32±0.76 | 4.70±0.84 | 7.72±1.96 | 0.0001 | 0.021 | 0.0002 |
| (5.31-7.67) | (2.93-6.29) | (4.49-10.53) | ||||
| Conformity index | 0.58±0.04 | 0.60±0.05 | 0.75±0.03 | 0.0003 | 0.00008 | 0.00008 |
| (0.50-0.67) | (0.49-0.71) | (0.68-0.83) | ||||
SD, standard deviation; WB-CRT, wedge-based conformal radiotherapy; FIF-IMRT, field-in-field intensity modulated radiotherapy; IMRT, intensity modulated radiotherapy; cGy, centigray.
Dosimetric Parameters for Organs at Risk with Three Planning Techniques
| Mean±SD (minimum-maximum) | ||||||
|---|---|---|---|---|---|---|
| WB-CRT | FIF-IMRT | IMRT | WB-CRT vs. FIF-IMRT | WB-CRT vs. IMRT | FIF-MRT vs. IMRT | |
| Right kidney Dmean (cGy) | 1,095.5±542.05 | 1,086.45±540.94 | 1,196.9±278.77 | 0.211 | 0.108 | 0.117 |
| (219-2265) | (218-2189) | (538-1764) | ||||
| Right kidney V20 (%) | 28.12±17.50 | 28.05±17.67 | 17.54±6.98 | 0.396 | 0.005 | 0.006 |
| (1.32-64.70) | (1.32-64.32) | (7.37-34.90) | ||||
| Left Kidney Dmean (cGy) | 998.15±441.77 | 1,000.90±441.45 | 1,334.20±230.14 | 0.059 | 0.001 | 0.001 |
| (199-1863) | (202-1876) | (840-1689) | ||||
| Left Kidney V20 (%) | 27.04±15.62 | 27.15±15.69 | 18.82±5.58 | 0.244 | 0.023 | 0.028 |
| (1.08-57.90) | (1.08-58.10) | (8.27-29.00) | ||||
| Liver Dmean (cGy) | 2,589.55±303.98 | 2,556.20±292.86 | 2,244.40±223.77 | 0.0002 | 0.0001 | 0.0001 |
| (2,057-3024) | (2,035-2955) | (1,951-2635) | ||||
| Liver V30 (%) | 35.07±11.48 | 33.77±10.82 | 23.75±4.99 | 0.005 | 0.0001 | 0.0002 |
| (16.00-64.08) | (16.00-60.95) | (17.50-32.93) | ||||
| Spinal cord Dmax (cGy) | 2,927.55±575.21 | 2,956.30±566.38 | 3,758.70±337.06 | 0.057 | 0.001 | 0.001 |
| (2,422-4652) | (2,424-4631) | (3,144-4333) | ||||
SD, standard deviation; WB-CRT, wedge-based conformal radiotherapy; FIF-IMRT, field-in-field intensity modulated radiotherapy; IMRT, intensity modulated radiotherapy; Dmean, mean dose; Dmax, maximum dose; V20, the volume receiving greater than or equal to 20 Gy of the prescribed dose; V30, the volume receiving greater than or equal to 30 Gy of the prescribed dose; cGy, centigray
Overdose Rates (Greater than Dose Constraints) for Organs at Risk with Each Technique
| WB-CRT | FIF-IMRT | IMRT | |
|---|---|---|---|
| n (%) | n (%) | n (%) | |
| Right kidney Dmean | 3 (15) | 3 (15) | 0 (0) |
| Right kidney V20 | 7 (35) | 7 (35) | 1 (5) |
| Left kidney Dmean | 1 (5) | 1 (5) | 0 (0) |
| Left kidney V20 | 6 (30) | 7 (35) | 0 (0) |
| Liver Dmean | 14 (70) | 13 (65) | 4 (20) |
| Liver V30 | 14 (70) | 13 (65) | 3 (15) |
| Spinal cord Dmax | 1 (5) | 1 (5) | 0 (0) |
WB-CRT, wedge-based conformal radiotherapy; FIF-IMRT, field-in-field intensity modulated radiotherapy; IMRT, intensity modulated radiotherapy; Dmean, mean dose; Dmax, maximum dose; V20, the volume receiving greater than or equal to 20 Gy of the prescribed dose; V30, the volume receiving greater than or equal to 30 Gy of the prescribed dose; n, number of patients