A Simple Cable Method for Intraoperative Limb Alignment Assessment.

C K Chan1, R Shanmugam1.   

Abstract

Entities:  

Year:  2016        PMID: 28435552      PMCID: PMC5333708     

Source DB:  PubMed          Journal:  Malays Orthop J        ISSN: 1985-2533


× No keyword cloud information.
Achieving good mechanical axis of the lower extremity is of paramount importance in orthopaedic surgery, not only during joint replacement but also in fracture fixation and deformity correction. However, it is difficult to determine the mechanical axis during surgery. Many techniques have been proposed in clinical practice such as visual inspection, cable method, alignment rods, the axis board[1], panoramic radiograph images[2] and navigation systems. Among all, cable method is proven to be one of the practical and uncomplicated way yet valid enough to control the mechanical axis intra-operatively[3]. The Limb Lengthening and Reconstruction Surgery (LLRS) unit of University Malaya Medical Centre devised a cable method to assess the mechanical axis alignment intraoperatively. Our apparatus includes a 1.5mm stainless steel cable, 2 aluminum hooks, 2 wooden bars, weight holder and weight (we used a 500 mls saline drip bottle in this setting) (Fig 1a). Aluminum plate is chosen as it is malleable and easily machined to desirable shapes. One end of the plate is bent so that it can be hooked onto the operating table side bars (Fig 1b). On the other end a hole with extension slot is created for attachment of the wooden bar (Fig 1a). Around the middle of the plate, we use a bolt and nut to serve as stopper to prevent tilting movement due to the pulling effect of the cable (Fig 1b).
Fig. 1a

This photograph shows the main components of limb alignment device.

Fig. 1b

This drawing shows how the alignment device was placed onto the side bar of operating table.

This photograph shows the main components of limb alignment device. This drawing shows how the alignment device was placed onto the side bar of operating table. Our apparatus can be installed under the operating table before surgery, and the position of the cable can be further adjusted during surgery. The proximal and distal end of the cable is positioned at the centre of the femoral head and ankle respectively and confirmed by intensifier imaging (Figure 2a-c). This simple cable method is cost effectiveness, simple to set up and easy to reproduce. We believe this apparatus is of substantial value in checking the coronal plane alignment during orthopaedic procedures of the lower limbs.
Fig. 2a

Proximal wire was positioned at the centre of femoral head.

Fig. 2c

At the knee level, position of he wire indicated medial deviation of the lower limb mechanical axis.

Proximal wire was positioned at the centre of femoral head. Distal wire was adjusted until the wire passed the centre of ankle / talus. At the knee level, position of he wire indicated medial deviation of the lower limb mechanical axis.
  3 in total

1.  Long bone X-ray image stitching using Camera Augmented Mobile C-arm.

Authors:  Lejing Wang; Joerg Traub; Sandro Michael Heining; Selim Benhimane; Ekkehard Euler; Rainer Graumann; Nassir Navab
Journal:  Med Image Comput Comput Assist Interv       Date:  2008

2.  A cadaver study comparing intraoperative methods to analyze lower limb alignment.

Authors:  Nael Hawi; Emmanouil Liodakis; Eduardo M Suero; Rupert Meller; Musa Citak; Christian Krettek
Journal:  Skeletal Radiol       Date:  2014-08-12       Impact factor: 2.199

3.  The axis-board: an alternative to the cable technique for intraoperative assessment of lower limb alignment.

Authors:  E Liodakis; M Kenawey; E Liodaki; P Mommsen; C Krettek; S Hankemeier
Journal:  Technol Health Care       Date:  2010       Impact factor: 1.285

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.