Literature DB >> 28432805

Pitfalls in electrogram interpretation: Subcutaneous cardioverter defibrillator malfunction in Brugada syndrome.

Giovanni Morani1, Bruna Bolzan1, Luca Tomasi1, Ruggero Tomei1, Corrado Vassanelli1.   

Abstract

A patient with Brugada syndrome implanted with subcutaneous implantable cardioverter defibrillator (S-ICD) had oversensing episodes treated with S-ICD shocks. Comparable artifacts were not evocable with S-ICD pocket manipulation. The fluoroscopy excluded S-ICD macroscopic damage. The device extraction revealed undamaged pulse generator and connector, but the lead was inappropriately tunneled under the sixth rib. Then the S-ICD malfunction was due to lead microscopic damage caused by the lead rubbing the rib surface.
© 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  cardioverter defibrillator malfunction; electrogram artefacts; subcutaneous cardioverter defibrillator; subcutaneous cardioverter defibrillator implantation technique

Mesh:

Year:  2017        PMID: 28432805     DOI: 10.1111/pace.13091

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

Review 1.  [Management of inappropriate shocks/T-wave-oversensing in S-ICD®-patients].

Authors:  Robert Larbig; Markus Bettin; Lukas J Motloch; Alicia Fischer; Niklas Bode; Gerrit Frommeyer; Florian Reinke; Andreas Loeher; Lars Eckardt; Julia Köbe
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2018-02-12

2.  Impact of a novel algorithm designed to reduce T-wave oversensing with the subcutaneous defibrillator in a patient with type I Brugada electrocardiogram.

Authors:  Serge Boveda; Marijke C Laarakker; Christèle Cardin; Jean-Paul Albenque
Journal:  HeartRhythm Case Rep       Date:  2017-11-21