Additional electrodes on the Quartet™ LV lead provide more programmable pacing options than bipolar and tripolar equivalents.

David O'Donnell1, Johannes Sperzel2, Bernard Thibault3, Christopher A Rinaldi4, Carlo Pappone5, Klaus-Jürgen Gutleben6, Christopher Leclercq7, Hedi Razavi8, Kyungmoo Ryu8, Luke C Mcspadden8, Avi Fischer8, Gery Tomassoni9.   

Abstract

AIMS: The aim of this study was to evaluate any benefits to the number of viable pacing vectors and maximal spatial coverage with quadripolar left ventricular (LV) leads when compared with tripolar and bipolar equivalents in patients receiving cardiac resynchronization therapy (CRT). METHODS AND
RESULTS: A meta-analysis of five previously published clinical trials involving the Quartet™ LV lead (St Jude Medical, St Paul, MN, USA) was performed to evaluate the number of viable pacing vectors defined as capture thresholds ≤2.5 V and no phrenic nerve stimulation and maximal spatial coverage of viable vectors in CRT patients at pre-discharge (n = 370) and first follow-up (n = 355). Bipolar and tripolar lead configurations were modelled by systematic elimination of two and one electrode(s), respectively, from the Quartet lead. The Quartet lead with its four pacing electrodes exhibited the greatest number of pacing vectors per patient when compared with the best bipolar and the best tripolar modelled equivalents. Similarly, the Quartet lead provided the highest spatial coverage in terms of the distance between two furthest viable pacing cathodes when compared with the best bipolar and the best tripolar configurations (P < 0.05). Among the three modelled bipolar configurations, the lead configuration with the two most distal electrodes resulted in the highest number of viable pacing vectors. Among the four modelled tripolar configurations, elimination of the second proximal electrode (M3) resulted in the highest number of viable pacing options per patient. There were no significant differences observed between pre-discharge and first follow-up analyses.
CONCLUSION: The Quartet lead with its four electrodes and the capability to pace from four anatomical locations provided the highest number of viable pacing vectors at pre-discharge and first follow-up visits, providing more flexibility in device programming and enabling continuation of CRT in more patients when compared with bipolar and tripolar equivalents. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2016. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  Cardiac resynchronization therapy (CRT); Device programming; Left ventricular leads; Quadripolar

Mesh:

Year:  2017        PMID: 28431058     DOI: 10.1093/europace/euw039

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  4 in total

Review 1.  Updates in Cardiac Resynchronization Therapy for Chronic Heart Failure: Review of Multisite Pacing.

Authors:  Antonios P Antoniadis; Ben Sieniewicz; Justin Gould; Bradley Porter; Jessica Webb; Simon Claridge; Jonathan M Behar; Christopher Aldo Rinaldi
Journal:  Curr Heart Fail Rep       Date:  2017-10

2.  Short-Term Availability of Viable Left Ventricular Pacing Sites with Quartet™ Quadripolar Leads.

Authors:  Min Gu; Wei Hua; Xiao-Han Fan; Li-Gang Ding; Jing Wang; Hong-Xia Niu; Cong Xu; Han Jin; Shu Zhang
Journal:  Med Sci Monit       Date:  2017-02-11

3.  Multi-site multi-polar left ventricular pacing through persistent left superior vena cava in tricuspid valve disease.

Authors:  Ernest W Lau
Journal:  Indian Pacing Electrophysiol J       Date:  2017-05-30

4.  Electrical remodelling and response following cardiac resynchronization therapy: A novel analysis of intracardiac electrogram using a quadripolar lead.

Authors:  Liam Toner; Darragh Flannery; Hariharan Sugumar; Michelle Ord; Tina Lin; David O'Donnell
Journal:  J Arrhythm       Date:  2018-04-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.