Abrupt disruption of remote monitoring transmission as an indicator of safe backup mode.

Yoichi Ajiro, Fumiaki Mori, Kazunori Iwade.   

Abstract

Entities:  

Mesh:

Year:  2017        PMID: 28431052      PMCID: PMC5400085          DOI: 10.1093/europace/euw203

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


× No keyword cloud information.
A 78-year-old man with sick sinus syndrome and accompanying persistent atrial tachyarrhythmia had undergone dual-chamber pacemaker (Evia DR-T™ Biotronik Inc., Germany) implantation with bipolar screw-in atrial and ventricular leads (Siello JT53™ and Siello S60™, Biotronik Inc.). Measured pacemaker and lead function values were within acceptable limits. Remote monitoring (Home Monitoring™ Biotronik Inc.) was used in addition to regular check-up visits for his out-of-hospital management. Data transmission via remote monitoring was favourable, with all parameters showing normal results. Twenty-two months after implantation, data transmission became disrupted. A 12-lead electrocardiography at an emergency pacemaker clinic indicated unipolar ventricular pacing (Figure). On interrogation, the pacemaker was found to be in the safe backup mode; the remote monitoring setting had been switched from ON to OFF, and pacemaker memories had been erased completely. Thorough inspection of the patient's surroundings did not identify the cause of the in-circuit excess current that resulted in the pacemaker setting changing to the safe backup mode. Pacemaker settings were reprogrammed to the former settings. The patient's clinical course was stable for 12 months thereafter. To our knowledge, this is the first case of abrupt disruption of remote monitoring transmission indicating a change to the safe backup mode. The full-length version of this report can be viewed at: http://www.escardio.org/Guidelines-&-Education/E-learning/Clinical-cases/Electrophysiology/EP-Case-Reports.
  1 in total

1.  Cause of the "power-on reset" phenomenon other than electric magnetic interference in a patient with a pacemaker.

Authors:  Kyoichiro Yazaki; Masahiro Watarai; Mitsuru Kahata; Asako Kumagai; Koji Inoue; Hiroshi Koganei; Kenji Enta; Masato Otsuka; Yasuhiro Ishii
Journal:  Indian Pacing Electrophysiol J       Date:  2018-02-22
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.