J P Prasad1, Y Madhu1, Arushi Mandhan1, Payal Garg1, Anu Prakash1, Varsha Singh1, Daud Ali1, Prdeep Kumar1, Surinder Singh1, G R Soni1, Varun Singh1.
Abstract
Current study is conducted to evaluate method verification of two locally available kits manufactured by DSI & BIORAD for quantitative estimation of Hepatitis B virus antibodies in human normal immunoglobulin by using International standard of National Institute of Biological Standards and Control. Four analyst perform five sets of test in duplicate analysing accuracy, precision, and limit of detection, sensitivity and specificity. Our results suggest that both DSI and BIORAD kits fulfil the validation criteria and are sensitive to detect up to 10 mIU concentration precisely and accurately. DSI kit is more precise at concentration 100 mIU and economically 4-5 times cheaper in local market; on the other hand, BIORAD kits provide larger detection range up to 1000 mIU.Entities:
Keywords: Accuracy; Hepatitis B virus; Immunoglobulin; Limit of detection; Precision; Sensitivity; Specificity
Year: 2016 PMID: 28428703 PMCID: PMC5382075 DOI: 10.1007/s12291-016-0590-9
Source DB: PubMed Journal: Indian J Clin Biochem ISSN: 0970-1915