In vitro fatigue tests and in silico finite element analysis of dental implants with different fixture/abutment joint types using computer-aided design models.

Satoshi Yamaguchi1, Yasufumi Yamanishi2, Lucas S Machado3, Shuji Matsumoto4, Nick Tovar3, Paulo G Coelho3, Van P Thompson5, Satoshi Imazato4.   

Abstract

PURPOSE: The aim of this study was to evaluate fatigue resistance of dental fixtures with two different fixture-abutment connections by in vitro fatigue testing and in silico three-dimensional finite element analysis (3D FEA) using original computer-aided design (CAD) models.
METHODS: Dental implant fixtures with external connection (EX) or internal connection (IN) abutments were fabricated from original CAD models using grade IV titanium and step-stress accelerated life testing was performed. Fatigue cycles and loads were assessed by Weibull analysis, and fatigue cracking was observed by micro-computed tomography and a stereomicroscope with high dynamic range software. Using the same CAD models, displacement vectors of implant components were also analyzed by 3D FEA. Angles of the fractured line occurring at fixture platforms in vitro and of displacement vectors corresponding to the fractured line in silico were compared by two-way ANOVA.
RESULTS: Fatigue testing showed significantly greater reliability for IN than EX (p<0.001). Fatigue crack initiation was primarily observed at implant fixture platforms. FEA demonstrated that crack lines of both implant systems in vitro were observed in the same direction as displacement vectors of the implant fixtures in silico.
CONCLUSIONS: In silico displacement vectors in the implant fixture are insightful for geometric development of dental implants to reduce complex interactions leading to fatigue failure.
Copyright © 2017 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bioengineering; Dental implant-abutment design; Fatigue test; Finite element analysis; Step-stress accelerated life tests

Mesh:

Substances:

Year:  2017        PMID: 28427837     DOI: 10.1016/j.jpor.2017.03.006

Source DB:  PubMed          Journal:  J Prosthodont Res        ISSN: 1883-1958            Impact factor:   4.642


  6 in total

1.  Complete mechanical characterization of an external hexagonal implant connection: in vitro study, 3D FEM, and probabilistic fatigue.

Authors:  María Prados-Privado; Sérgio A Gehrke; Rosa Rojo; Juan Carlos Prados-Frutos
Journal:  Med Biol Eng Comput       Date:  2018-06-11       Impact factor: 2.602

2.  Assessment of automatic segmentation of teeth using a watershed-based method.

Authors:  Antoine Galibourg; Jean Dumoncel; Norbert Telmon; Adèle Calvet; Jérôme Michetti; Delphine Maret
Journal:  Dentomaxillofac Radiol       Date:  2017-11-01       Impact factor: 2.419

3.  Biomechanical evaluations of the long-term stability of dental implant using finite element modeling method: a systematic review.

Authors:  Seyed Aref Hosseini-Faradonbeh; Hamid Reza Katoozian
Journal:  J Adv Prosthodont       Date:  2022-06-27       Impact factor: 1.989

4.  Validation of finite-element simulations with synchrotron radiography - A descriptive study of micromechanics in two-piece dental implants.

Authors:  Wolfram Wiest; Alexander Rack; Simon Zabler; Alex Schaer; Michael Swain; Katja Nelson
Journal:  Heliyon       Date:  2018-02-08

5.  Classification of the journal category "oral surgery" in the Scopus and the Science Citation Index Expanded: flaws and suggestions.

Authors:  Seong-Gon Kim
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2019-08-28

6.  Mechanical Assessment of Fatigue Characteristics between Single- and Multi-Directional Cyclic Loading Modes on a Dental Implant System.

Authors:  Won Hyeon Kim; Eun Sung Song; Kyung Won Ju; Dohyung Lim; Dong-Wook Han; Tae-Gon Jung; Yong-Hoon Jeong; Jong-Ho Lee; Bongju Kim
Journal:  Materials (Basel)       Date:  2020-03-27       Impact factor: 3.623

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.