Johannes Kalbhenn1, Sven Maier2, Sebastian Heinrich3, Nils Schallner3.
Abstract
We report a repositioning maneuver of a dislocated Avalon Elite dual lumen catheter during ongoing veno-venous ECMO support. The inferior tip of the catheter dislocated into a liver vein, which was accompanied by a dramatic decrease in pump flow. After standard repositioning maneuvers under transthoracic echocardiographic guidance had failed, a special guiding sheath was inserted into the main lumen through a Y-connector with a valve. Over this valve, a stiff wire could be placed into the inferior vena cava to help guiding the catheter back into the correct position.Keywords: ARDS; Avalon; Cannula dislocation; ECMO; Echocardiography; Transthoracic