Literature DB >> 28421237

Radiographic interpretation of carpometacarpal arthroplasty: correlation between radiographic loosening and clinical outcome.

Laurence Stillwater1, Brett Memauri2, Imran Ratanshi2, Avinash Islur2, Thelina Amaratunga2.   

Abstract

OBJECTIVE: To assess the radiographic appearance and interpretation of loosening in patients following first carpometacarpal joint pyrohemisphere arthroplasty.
MATERIALS AND METHODS: A retrospective review over a 6-year period was performed identifying patients with first carpometacarpal pyrohemisphere implants. All postoperative radiographs were reviewed and compared with clinical information. Loosening was defined as greater than 1 mm of periprosthetic lucency or increasing lucency on serial studies. Adverse clinical outcome was defined as infection, complex regional pain syndrome, subluxation or an inability to return to activities of daily living. Statistical analysis included Cohen's Kappa coefficient to measure inter-reader agreement for radiographic interpretation and the agreement between radiographic loosening and an undesired or adverse clinical outcome.
RESULTS: Thirty-one implants in 26 patients were reviewed. A total of 73 postoperative radiographs were reviewed. The mean age of the study subjects was 58 years. The mean radiographic follow-up was 13 months (range: 1-56). All arthroplasties were performed for osteoarthritis. Only one repeat surgery was performed. There was good agreement amongst readers with regard to radiographic interpretation, but the strength of agreement between radiographic loosening and adverse clinical outcome was poor. The sensitivity of radiography in predicting an adverse clinical outcome was 63%, specificity 65%, positive predictive value 39%, and negative predictive value 83%.
CONCLUSION: Although radiography may be useful in the correct clinical context, it should not be utilized as the sole predictor of adverse clinical outcomes following carpometacarpal arthroplasty.

Entities:  

Keywords:  Arthroplasty; Arthroplasty loosening; CMC; CMC joint; Carpometacarpal; Carpometacarpal arthroplasty; Carpometacarpal joint; Hand; Implant; Implant loosening; Loosening; Osteoarthritis; Pyrocarbon; Pyrocarbon implant

Mesh:

Year:  2017        PMID: 28421237     DOI: 10.1007/s00256-017-2648-z

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  24 in total

1.  The Nicolle finger joint prosthesis: a reappraisal.

Authors:  S K Varma; T M Milward
Journal:  J Hand Surg Br       Date:  1991-05

2.  Problematic bone fixation with pyrocarbon implants in proximal interphalangeal joint replacement: short-term results.

Authors:  D B Herren; S Schindele; J Goldhahn; B R Simmen
Journal:  J Hand Surg Br       Date:  2006-10-12

3.  Long-term results of proximal interphalangeal joint resection arthroplasties with a silicone implant.

Authors:  F Iselin; E Conti
Journal:  J Hand Surg Am       Date:  1995-05       Impact factor: 2.230

4.  Pyrolytic carbon proximal interphalangeal joint arthroplasty: results with minimum two-year follow-up evaluation.

Authors:  César J Bravo; Marco Rizzo; Kirsten B Hormel; Robert D Beckenbaugh
Journal:  J Hand Surg Am       Date:  2007-01       Impact factor: 2.230

5.  Long-term follow-up of pyrolytic carbon metacarpophalangeal implants.

Authors:  S D Cook; R D Beckenbaugh; J Redondo; L S Popich; J J Klawitter; R L Linscheid
Journal:  J Bone Joint Surg Am       Date:  1999-05       Impact factor: 5.284

6.  Short-term results of the Ascension pyrolytic carbon metacarpophalangeal joint replacement arthroplasty for osteoarthritis.

Authors:  V A Nuñez; N D Citron
Journal:  Chir Main       Date:  2005 Jun-Aug

7.  Pyrolytic carbon arthroplasty for posttraumatic arthritis of the proximal interphalangeal joint.

Authors:  Ryan M Nunley; Martin I Boyer; Charles A Goldfarb
Journal:  J Hand Surg Am       Date:  2006-11       Impact factor: 2.230

8.  Metacarpophalangeal joint arthroplasty with a metallic hinged prosthesis.

Authors:  W F Blair; D G Shurr; J A Buckwalter
Journal:  Clin Orthop Relat Res       Date:  1984-04       Impact factor: 4.176

9.  Arthroplasty of the metacarpophalangeal joints with use of silicone-rubber implants in patients who have rheumatoid arthritis. Long-term results.

Authors:  D Kirschenbaum; L H Schneider; D C Adams; R P Cody
Journal:  J Bone Joint Surg Am       Date:  1993-01       Impact factor: 5.284

10.  Resurfacing arthroplasty versus silicone arthroplasty for proximal interphalangeal joint osteoarthritis.

Authors:  Barton R Branam; Harrison G Tuttle; Peter J Stern; Linda Levin
Journal:  J Hand Surg Am       Date:  2007 Jul-Aug       Impact factor: 2.230

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.