Literature DB >> 28415470

Spatial stress distribution analysis by thermoelastic stress measurement and evaluation of effect of stress concentration on durability of various orthopedic implant devices.

Yoshimitsu Okazaki1, Daisuke Ishii2, Atsushi Ogawa2.   

Abstract

Toward the development of highly durable devices, we investigated the effect of the thermoelastic constants of implantable raw metals and the surface stress distribution on the durability of various types of implant device by thermoelastic stress measurement and by evaluating the effect of the stress concentration. Surface stress was dynamically calculated from the bending moment, and the modulus of a section of a device was found to be consistent with the surface stress obtained by thermoelastic stress measurement. The durability limits of various types of bone plate and compression hip screw (CHS) calculated from maximum load vs number of cycles data (L-N data) were close to the notch fatigue strength of the raw material. The concentration factor of an artificial hip stem surface was estimated by comparing the L-N data of the stem and the S-N curve of the raw material. The dynamic analysis of durability by thermoelastic stress measurement is useful for selecting the worst case (a product deteriorating to the most severe state) in medical device design.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Bending moment; Durability; Fatigue property; Metallic implant; Raw material; Stress distribution; Thermoelastic stress measurement

Mesh:

Year:  2017        PMID: 28415470     DOI: 10.1016/j.msec.2017.02.014

Source DB:  PubMed          Journal:  Mater Sci Eng C Mater Biol Appl        ISSN: 0928-4931            Impact factor:   7.328


  2 in total

1.  A Hybrid Uniplanar Pedicle Screw System with a New Intermediate Screw for Minimally Invasive Spinal Fixation: A Finite Element Analysis.

Authors:  Jia Li; Li-Cheng Zhang; Jiantao Li; Hao Zhang; Jing-Xin Zhao; Wei Zhang
Journal:  Biomed Res Int       Date:  2020-11-18       Impact factor: 3.411

2.  Strength⁻Durability Correlation of Osteosynthesis Devices Made by 3D Layer Manufacturing.

Authors:  Yoshimitsu Okazaki; Emiko Gotoh; Jun Mori
Journal:  Materials (Basel)       Date:  2019-01-31       Impact factor: 3.623

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.