Characterization of the surface physico-chemistry of plasticized PVC used in blood bag and infusion tubing.

H Al Salloum1, J Saunier2, A Dazzi3, J Vigneron4, A Etcheberry4, C Marlière5, C Aymes-Chodur1, J M Herry6, M Bernard1, E Jubeli1, N Yagoubi1.   

Abstract

Commercial infusion tubing and blood storage devices (tubing, blood and platelets bags) made of plasticized PVC were analyzed by spectroscopic, chromatographic and microscopic techniques in order to identify and quantify the additives added to the polymer (lubricants, thermal stabilizers, plasticizers) and to put into evidence their blooming onto the surface of the devices. For all the samples, deposits were observed on the surface but with different kinds of morphologies. Ethylene bis amide lubricant and metallic stearate stabilizers were implicated in the formation of these layers. In contact with aqueous media, these insoluble deposits were damaged, suggesting a possible particulate contamination of the infused solutions.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Blood bag; Blooming; Infusion; Lubricant; Metallic stearate; PVC; Particulate contamination

Mesh:

Substances:

Year:  2017        PMID: 28415468     DOI: 10.1016/j.msec.2017.02.057

Source DB:  PubMed          Journal:  Mater Sci Eng C Mater Biol Appl        ISSN: 0928-4931            Impact factor:   7.328


  2 in total

1.  DEHP Nanodroplets Leached From Polyvinyl Chloride IV Bags Promote Aggregation of IVIG and Activate Complement in Human Serum.

Authors:  Jared R Snell; Connor R Monticello; Cheng Her; Emma L Ross; Ashley A Frazer-Abel; John F Carpenter; Theodore W Randolph
Journal:  J Pharm Sci       Date:  2019-06-21       Impact factor: 3.534

2.  Effects of flow rate on the migration of different plasticizers from PVC infusion medical devices.

Authors:  Lise Bernard; Teuta Eljezi; Hélène Clauson; Céline Lambert; Yassine Bouattour; Philip Chennell; Bruno Pereira; Valérie Sautou
Journal:  PLoS One       Date:  2018-02-23       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.