Literature DB >> 28414352

Infrared absorption spectrometer for the determination of temperature and species profiles in an entrained flow gasifier.

Patrick Nau, Peter Kutne, Georg Eckel, Wolfgang Meier, Christian Hotz, Sabine Fleck.   

Abstract

An absorption spectrometer utilizing a tunable distributed feedback diode laser at 2.3 μm and an interband cascade laser at 3.1 μm has been developed to measure temperature and concentrations of CO, CH<sub>4</sub>, C<sub>2</sub>H<sub>2</sub>, and H<sub>2</sub>O under gasification conditions. A wavelength division multiplexing approach using a single ZrF<sub>4</sub>-fiber was used to measure both wavelength regions simultaneously. The performance of the spectrometer has been tested in laminar flat flames and a heated cell and then applied for measurements at an atmospheric entrained flow gasifier (REGA). A water-cooled optical probe was used to provide optical access at two measurement positions. By moving the burner, axial profiles of temperature and species concentration could be obtained. These profiles were compared with numerical simulations and can be used to validate the simulation.

Entities:  

Year:  2017        PMID: 28414352     DOI: 10.1364/AO.56.002982

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  A Compact Fiber-Coupled NIR/MIR Laser Absorption Instrument for the Simultaneous Measurement of Gas-Phase Temperature and CO, CO2, and H2O Concentration.

Authors:  Lin Shi; Torsten Endres; Jay B Jeffries; Thomas Dreier; Christof Schulz
Journal:  Sensors (Basel)       Date:  2022-02-08       Impact factor: 3.576

  1 in total