| Literature DB >> 28412961 |
Fei Liang1, Li Hongxin2, Hai-Zhou Zhang1, Guo Wenbin3, Cheng-Wei Zou1, Zeeshan Farhaj1.
Abstract
BACKGROUND: Device closure of a wide-spaced multi-hole PmVSD is difficult to succeed in percutaneous approach. This study is to evaluate the feasibility, safety and efficacy of perventricular device closure of wide-spaced multi-hole PmVSD using a double-device implanting technique.Entities:
Keywords: Device closure; Echocardiography; Perimembranous ventricular septal defect; Perventricular
Mesh:
Year: 2017 PMID: 28412961 PMCID: PMC5392910 DOI: 10.1186/s13019-017-0585-5
Source DB: PubMed Journal: J Cardiothorac Surg ISSN: 1749-8090 Impact factor: 1.637
Baseline clinical characteristic data and procedural data
| No | Sex | Age (y) | Weight (kg) | LDVSD (mm) | Number of holes | Size of holes (mm) | MD (mm) | 1st device (mm) | 2nd device (mm) | ICMT (min) | Procedure time (min) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | M | 39 | 80 | 19 | 4 | 7.0,2.0,2.5,1.5 | 7.2 | 9,sc | 5,wc | 68 | 135 |
| 2 | M | 1 | 9 | 8 | 2 | 2.6,2.0 | 7.0 | 4,sc | 4,wc | 45 | 75 |
| 3 | M | 1 | 12 | 9 | 2 | 2.7,2.5 | 5.3 | 4,sc | 4,sc | 50 | 88 |
| 4 | M | 3 | 18 | 12 | 2 | 6.0,2.5 | 9.0 | 8,sc | 4,sc | 47 | 115 |
| 5 | M | 3 | 16 | 14 | 4 | 5.6,2.5,1.5,1.5 | 10.0 | 8,sc | 5,wc | 48 | 100 |
| 6 | M | 20 | 56 | 18 | 2 | 3.2,2.5 | 6.5 | 6,sc | 4,sc | 90 | 160 |
| 7 | F | 3 | 14 | 12 | 2 | 2.5,2.5 | 5.5 | 4,sc | 4,wc | 45 | 84 |
| 8 | M | 19 | 64 | 15 | 2 | 3.5,2.5 | 6.0 | 6,sc | 4,sc | 65 | 108 |
| 9 | M | 3 | 13 | 11 | 2 | 2.8,2.2 | 6.1 | 4,sc | 4,sc | 53 | 100 |
| 10 | M | 6 | 19 | 13 | 3 | 3.2,2.0,1.5 | 5.0 | 7,wc | 4,sc | 40 | 99 |
| 11 | F | 1 | 7 | 10 | 3 | 2.5,2.0,1.5 | 5.3 | 5,sc | 4,sc | 45 | 120 |
| 12 | F | 1 | 12 | 9 | 3 | 2.5,2.0,1.5 | 5.5 | 5,wc | 4,sc | 33 | 80 |
| 13 | F | 8 | 25 | 17 | 2 | 3.8,2.5 | 6.5 | 6,sc | 4,sc | 60 | 105 |
| 14 | F | 5 | 28 | 10 | 2 | 3.0,2.3 | 5.2 | 5,sc | 4,sc | 50 | 98 |
| 15 | F | 13 | 54 | 15 | 2 | 3.5,2.5 | 5.8 | 6,sc | 4,sc | 62 | 108 |
| 16 | M | 4 | 19 | 9 | 2 | 3.0,2.5 | 5.3 | 5,sc | 4,sc | 38 | 70 |
LDVSD Left-sided diameter of ventricular septal defect, MD maximum distance between the holes, sc concentric short-edged occluder, wc concentric wide-edged occluder, ICMT intracardiac manipulation time
Fig. 1a The direct delivery system consists of a short delivery sheath and a delivery cable. (magnification of inset 2.5 ×) b The probe-assisted delivery system consists of a hollow probe, a flexible guidewire, a delivery sheath, a loading sheath and a delivery cable
Fig. 2a The PmVSD had two holes, and the maximum distance between the holes was more than 5 mm; b The largest hole was occluded with the direct delivery system; c The residual hole was occluded with the probe-assisted delivery system; d The double devices were positioned properly, and there was no residual shunt
Fig. 3The chest radiography showed the arrangement and a good configuration of the two devices. a AP view; b Lateral view
Postoperative and follow-up results
| Variable | IADR | At discharge | 1 Month | 6 Months | 12 Months | 24 Months | 36 Months | 48 Months |
|---|---|---|---|---|---|---|---|---|
| Preexisting AR (%) | 2/16(13) | 1/16(6) | 0/16(0) | 0/16(0) | 0/16(0) | 0/11(0) | 0/7(0) | 0/3(0) |
| New AR (%) | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Preexisting TR (%) | 8/16(50) | 7/16(44) | 4/16(25) | 1/16(6) | 1/16(6) | 1/11(9) | 0/7(0) | 0/3(0) |
| New TR (%) | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Preexisting MR (%) | 5/16(31) | 2/16(13) | 2/16(13) | 1/16(6) | 0/16(0) | 0/11(0) | 0/7(0) | 0/3(0) |
| New MR (%) | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| New IRBBB (%) | – | 3/16(19) | 2/16(13) | 1/16(6) | 1/16(6) | 0/11(0) | 0/7(0) | 0/3(0) |
| Mild residual shunt (%) | 6/16(38) | 5/16(31) | 3/16(19) | 1/16(6) | 1/16(6) | 0/11(0) | 0/7(0) | 0/3(0) |
AR aortic regurgitation, TR tricuspid regurgitation, MR mitral regurgitation, IRBBB incomplete right bundle-branch block, IADR immediately after device release