| Literature DB >> 28412689 |
Abdul Sami Shaikh1, Wenlong Li2, Guiyan Yuan2, Meimei Gao2, Chunmei Geng2, Nan Guo2, Ruichen Guo2.
Abstract
Spikes in Lamotrigine concentrations levels and associated clinical toxicity may occur unpredictably. This study describes the development and validation of a simple, more rapid, highly sensitive and economical method for measuring Lamotrigine (LTG) concentration levels in human plasma using HPLC-UV and its clinical applications. Analyte from plasma was extracted with methanol (protein precipitation) and separated on the analytical column Diamonsil C18 (150mm×4.6mm, 5μm) Waters-Milford, MA, United States. Mixture of 0.1% Trifluoroacetate and Methanol used as mobile phase in a 59:41 volume/volume mixture with an isocratic flow rate of 1.5 ml/min and wavelength was adjusted to 260nm. Standard curve of lamotrigine showed good linearity over the range of 1.0-50μg/mL (r2=0.9961) and LLOQ was 1.0μg/ml. The Specificity, Recovery, Accuracy, Stability, Robustness and RSDs for both intraday and interday precision were within acceptable limits. The highly sensitive HPLC assay for determination of LTG in human plasma was demonstrated, validated and applied in Therapeutic Drug Monitoring (TDM) of sixty seven epilepsy patients who were using LTG. The proposed method can be easily applied in routine Therapeutic monitoring of LTG, Besides TDM, stated method can be also very useful for Bioequivalence studies, Pharmacovigilance and Pharmacokinetics studies.Entities:
Year: 2016 PMID: 28412689
Source DB: PubMed Journal: Pak J Pharm Sci ISSN: 1011-601X Impact factor: 0.684