The evaluation of the neutron dose equivalent in the two-bend maze.

Á Á Tóth1, B Petrović2, N Jovančević3, M Krmar3, L Rutonjski4, O Čudić4.   

Abstract

The purpose of this study was to explore the effect of the second bend of the maze, on the neutron dose equivalent, in the 15MV linear accelerator vault, with two bend maze. These two bends of the maze were covered by 32 points where the neutron dose equivalent was measured. There is one available method for estimation of the neutron dose equivalent at the entrance door of the two bend maze which was tested using the results of the measurements. The results of this study show that the neutron equivalent dose at the door of the two bend maze was reduced almost three orders of magnitude. The measured TVD in the first bend (closer to the inner maze entrance) is about 5m. The measured TVD result is close to the TVD values usually used in the proposed models for estimation of neutron dose equivalent at the entrance door of the single bend maze. The results also determined that the TVD in the second bend (next to the maze entrance door) is significantly lower than the TVD values found in the first maze bend.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Neutron dose; Tenth-value distance; Two-bend maze

Mesh:

Year:  2017        PMID: 28410680     DOI: 10.1016/j.ejmp.2017.03.017

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  The influence of shielding reinforcement in a vault with limited dimensions on the neutron dose equivalent in vicinity of medical electron linear accelerator.

Authors:  Ana Ivkovic; Dario Faj; Mladen Kasabasic; Marina Poje Sovilj; Ivana Krpan; Marina Grabar Branilovic; Hrvoje Brkic
Journal:  Radiol Oncol       Date:  2020-05-02       Impact factor: 2.991

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.