Comparison of Gravity Flow Rates Between ENFit and Legacy Feeding Tubes.

Manpreet S Mundi1, Wanda Duellman1, Lisa Epp1, Jacob Davidson2, Ryan T Hurt1,2,3,4.   

Abstract

BACKGROUND: Misconnections between enteral supplies and other access devices have led to significant morbidity and mortality. To reduce misconnections, a standard small-bore connector has been developed (International Organization for Standards 80369-8; ENFit). The full impact of transition to this connector is not known, however.
METHOD: Working with major manufacturers and Food and Drug Administration, we obtained ENFit and comparative legacy tubes of variable sizes (low-profile, 14F, 18F, 20F, and 24F balloon gastrostomies). Gravity enteral feeding was simulated with an empty bolus syringe attached to the feeding tube to be tested. The tube was clamped and filled to the 60-mL mark with liquid (water, Jevity 1 Cal, Isosource HN, Isosource 1.5 Cal, Two Cal HN, and Nourish). The clamp was released, and time for formula to leave the syringe was recorded.
RESULTS: There was no difference in flow rate between the aggregate legacy and ENFit tubes for the low-profile, 18F, and 20F sizes. The ENFit 14F tubes had a lower flow rate vs the legacy tubes, largely due to the low flow rates seen with the 1 ENFit tube. Similarly, 24F ENFit tubes with some formulas yielded lower flow rates as opposed to legacy.
CONCLUSION: Overall, for the low-profile, 18F, and 20F sizes, the ENFit tubes had similar flow rates when compared with the legacy tubes. For the 14F and 24F sizes, the flow rate of ENFit tubes was significantly lower, which could result in longer EN delivery for patients who are using these tubes to provide gravity feeding.
© 2017 American Society for Parenteral and Enteral Nutrition.

Entities:  

Keywords:  enteral nutrition; long-term care; nutrition; nutrition support practice; nutrition support teams

Mesh:

Year:  2017        PMID: 28406738     DOI: 10.1177/0148607117703959

Source DB:  PubMed          Journal:  JPEN J Parenter Enteral Nutr        ISSN: 0148-6071            Impact factor:   4.016


  1 in total

1.  Technical considerations for medical device manufacturers when designing gastrostomy tubes (G-tubes) using the new ISO 80369-3 connector.

Authors:  Suvajyoti Guha; Alexander Herman; Luke Herbertson; Mark J Antonino; Joshua S Silverstein; Jeffrey Cooper; Matthew R Myers
Journal:  PLoS One       Date:  2020-07-30       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.