IQ-SPECT for thallium-201 myocardial perfusion imaging: effect of normal databases on quantification.

Takahiro Konishi1, Kenichi Nakajima2, Koichi Okuda3, Hiroto Yoneyama1, Shinro Matsuo4, Takayuki Shibutani5, Masahisa Onoguchi5, Seigo Kinuya4.   

Abstract

OBJECTIVE: Although IQ-single-photon emission computed tomography (SPECT) provides rapid acquisition and attenuation-corrected images, the unique technology may create characteristic distribution different from the conventional imaging. This study aimed to compare the diagnostic performance of IQ-SPECT using Japanese normal databases (NDBs) with that of the conventional SPECT for thallium-201 (201Tl) myocardial perfusion imaging (MPI).
METHODS: A total of 36 patients underwent 1-day 201Tl adenosine stress-rest MPI. Images were acquired with IQ-SPECT at approximately one-quarter of the standard time of conventional SPECT. Projection data acquired with the IQ-SPECT system were reconstructed via an ordered subset conjugate gradient minimizer method with or without scatter and attenuation correction (SCAC). Projection data obtained using the conventional SPECT were reconstructed via a filtered back projection method without SCAC. The summed stress score (SSS) was calculated using NDBs created by the Japanese Society of Nuclear Medicine working group, and scores were compared between IQ-SPECT and conventional SPECT using the acquisition condition-matched NDBs. The diagnostic performance of the methods for the detection of coronary artery disease was also compared.
RESULTS: SSSs were 6.6 ± 8.2 for the conventional SPECT, 6.6 ± 9.4 for IQ-SPECT without SCAC, and 6.5 ± 9.7 for IQ-SPECT with SCAC (p = n.s. for each comparison). The SSS showed a strong positive correlation between conventional SPECT and IQ-SPECT (r = 0.921 and p < 0.0001), and the correlation between IQ-SPECT with and without SCAC was also good (r = 0.907 and p < 0.0001). Regarding diagnostic performance, the sensitivity, specificity, and accuracy were 80.8, 78.9, and 79.4%, respectively, for the conventional SPECT; 80.8, 80.3, and 82.0%, respectively, for IQ-SPECT without SCAC; and 88.5, 86.8, and 87.3%, respectively, for IQ-SPECT with SCAC, respectively. The area under the curve obtained via receiver operating characteristic analysis were 0.77, 0.80, and 0.86 for conventional SPECT, IQ-SPECT without SCAC, and IQ-SPECT with SCAC, respectively (p = n.s. for each comparison).
CONCLUSIONS: When appropriate NDBs were used, the diagnostic performance of 201Tl IQ-SPECT was comparable with that of the conventional system regardless of different characteristics of myocardial accumulation in the conventional system.

Entities:  

Keywords:  IQ-SPECT; Japanese normal databases; Myocardial perfusion imaging; Rapid acquisition

Mesh:

Substances:

Year:  2017        PMID: 28405882     DOI: 10.1007/s12149-017-1170-7

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  2 in total

Review 1.  IQ·SPECT technology and its clinical applications using multicenter normal databases.

Authors:  Kenichi Nakajima; Koichi Okuda; Mitsuru Momose; Shinro Matsuo; Chisato Kondo; Masayoshi Sarai; Takayuki Shibutani; Masahisa Onoguchi; Takeshi Shimizu; A Hans Vija
Journal:  Ann Nucl Med       Date:  2017-09-22       Impact factor: 2.668

2.  Diagnostic value of thallium-201 myocardial perfusion IQ-SPECT without and with computed tomography-based attenuation correction to predict clinically significant and insignificant fractional flow reserve: A single-center prospective study.

Authors:  Haruki Tanaka; Teruyuki Takahashi; Norihiko Ohashi; Koichi Tanaka; Takenori Okada; Yasuki Kihara
Journal:  Medicine (Baltimore)       Date:  2017-12       Impact factor: 1.817

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.