CLSI-Based Validation of Manufacturer-Derived Reference Intervals on the Cobas 8000 Platform.

Veronika Leitner-Ferenc1, Johanna Atamaniuk1, Sonja Jansen-Skoupy1, Brigitta Stöckelmeier1, Katharina Grohs1, Manuela Födinger1,2.   

Abstract

BACKGROUND: Reference intervals provided by diagnostic test manufacturers should be transferred to clinical laboratories after validation. Although protocols exist, laboratories rarely perform and report on results of validation studies.
METHODS: We validated reference intervals (RIs) of 87 analytes on a Cobas 8000 platform according to standards published by the Clinical and Laboratory Standards Institute (CLSI).
RESULTS: For 8 analytes, decision limits were provided in the package inserts. Among the 79 RIs subjected to transference validation, 8 were found not valid for transference, including lactate dehydrogenase (LDH) among women, and the following among both sexes: potassium, homocysteine, immunoglobulin E (IgE), free lambda light chain (FLC λ), C3 complement (C3c), folate, and 25-hydroxy vitamin D (25[(OH]D). For LDH, potassium, homocysteine, C3c, folate, and 25(OH)D, RIs or thresholds suitable for transference were available in the literature; however, this was not the case for IgE and FLC λ.
CONCLUSION: The present study demonstrates that validation of RIs provided in the manufacturer provided package inserts is indispensable. © American Society for Clinical Pathology, 2017. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com

Entities:  

Keywords:  analytical system; decision limit; guideline; reference values; test provider; transference

Mesh:

Year:  2017        PMID: 28398583     DOI: 10.1093/labmed/lmx020

Source DB:  PubMed          Journal:  Lab Med        ISSN: 0007-5027


  2 in total

1.  Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M.

Authors:  Mingxing Chen; Simeng Qin; Sitao Yang; Huaping Chen; Liuyi Lu; Xue Qin
Journal:  J Med Biochem       Date:  2022-07-29       Impact factor: 2.157

2.  Establishment and Verification of Sex- and Age-Specific Serum Electrolyte Reference Intervals in Healthy Han Children in Changchun, Northeastern China.

Authors:  Qi Zhou; Xin Li; Yanan Jia; Wenjia Guo; Baojie Guan; Jiancheng Xu
Journal:  Biomed Res Int       Date:  2019-11-26       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.