Merica Aralica1, Jasna Lenicek Krleza2.
Abstract
INTRODUCTION: Sweat test has a diagnostic role in evaluation of cystic fibrosis. Its performance includes sweat stimulation, collection and analysis. All listed may be sources of inconsistencies in everyday practice. The aim of this study was an evaluation of external quality assessment (EQA) of sweat chloride measurement including sweat test performance in medical biochemistry laboratories in Croatia.Entities:
Keywords: external quality assessment; survey; sweat testing
Mesh:
Year: 2017 PMID: 28392735 PMCID: PMC5382848 DOI: 10.11613/BM.2017.016
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Results of Croatian Centre for Quality Assessment in Laboratory Medicine (CROQALM) exercises in chloride concentrations of the sweat test scheme
| 1st (June 2015) | 8 | 53.0 ± 15.0 | 51.5 (36.3 – 128.0) | 28.9 | 0 |
| 2nd(September 2015) | 8 | 25.5 ± 7.4 | 23.9 (16.0 – 36.0) | 29.0 | 1 |
| 3rd (November 2015) | 8 | 45.4 ± 16.1 | 52.8 (22.5 – 64.6) | 35.3 | 3 |
| 1st (March 2016) | 8 | 29.7 ± 4.6 | 28.5 (24.8 – 37.0) | 15.5 | 0 |
| 2nd (June 2016) | 9 | 21.5 ± 3.2 | 19.9 (18.0 – 26.0) | 14.7 | 2 |
| N – Number of participating laboratories. SD - standard deviation. CV – coefficient of variation. | |||||
Questionnaire of sweat test performance in Croatian medical biochemistry laboratories.
| 10/10 | ||
| a) <50 | 2 | |
| b) 50-100 | 3 | |
| c) 100-300 | 3 | |
| d) >300 | 3 | |
| 10/10 | ||
| a) in a clinic | 4 | |
| b) in a laboratory | 6 | |
| 10/10 | ||
| a) pilocarpine iontophoresis | 10 | |
| b) wrapping a patient with plastic material | 0 | |
| 10/10 | ||
| a) laboratory technician and/or technologist | 5 | |
| b) medical biochemist | 1 | |
| c) clinical staff | 4 | |
| 10/10 | ||
| a) only written | 0 | |
| b) only verbal | 5 | |
| c) verbal and written | 2 | |
| d) by hospital web site | 1 | |
| e) no information provided | 2 | |
| 10/10 | ||
| a) yes | 2 | |
| b) no | 8 | |
| 10/10 | ||
| a) yes | 1 | |
| b) no | 9 | |
| 10/10 | ||
| a) 1 | 4 | |
| b) 2 | 5 | |
| c) > 3 | 0 | |
| 9/9 | ||
| a) stainless steel electrodes | 0 | |
| b) cooper electrodes | 3 | |
| c) electrodes of flexible size and curvature made by local manufacturer | 6 | |
| d) electrodes of unspecified origin | 0 | |
| 9/9 | ||
| a) < 30 min | 1 | |
| b) 30 min | 8 | |
| c) 30 - 40 min | 0 | |
| d) > 40 min | 0 | |
| 9/9 | ||
| a) 50 mg | 1 | |
| b) 75 mg | 5 | |
| c) 100 mg | 2 | |
| d) no criteria for acceptable sweat weight | 1 | |
| e) other, specify | 0 | |
| 9/9 | ||
| a) pre-weighed gauze | 7 | |
| b) pre-weighed filter paper | 2 | |
| c) polypropylene or glass tube | 0 | |
| 9/10 | ||
| a) keep testing and carry out chloride analysis | 1 | |
| b) stop testing and repeat sweat collection another time | 7 | |
| c) carry out bilateral sweat collection | 1 | |
| 9/9 | ||
| a) yes, specify | 5 | |
| b) no | 4 | |
| 10/10 | ||
| a) ≤ 10 | 0 | |
| b) 11 - 49 | 4 | |
| c) ≥ 50 | 5 | |
| d) no record | 1 | |
| 8/9 | ||
| a) negative < 30 mmol/L; intermediate 30 - 60 mmol/L; positive > 60 mmol/L | 3 | |
| b) negative < 60 mmol/L; intermediate 60 - 80 mmol/L; positive > 80 mmol/L | 3 | |
| c) other, specify | 2 | |
| 9/9 | ||
| a) negative < 40 mmol/L; intermediate 40 - 60 mmol/L; positive > 60 mmol/L | 5 | |
| b) negative < 60 mmol/L; intermediate 60 - 80 mmol/L; positive > 80 mmol/L | 1 | |
| c) other, specify | 3 | |
| 9/9 | ||
| a) yes, specify | 3 | |
| b) no | 6 | |
| N – number of responders. TN – total number of participating laboratories. | ||
Stratified responses of medical biochemistry laboratory in national questionnaire.
| d | a | a | c | c | b | b | b | c | b | c | a | b | a | A | c | b | a | a | ||
| c | b | a | a | b | a | a | NR | c | b | b | a | b | a | B | c | b | b | a | ||
| c | b | a | a | c | b | b | b | b | b | b | b | b | a | B | b | a | a | b | ||
| b | a | a | c | e | b | b | b | c | a | a | a | b | b | - | b | NR | c | b | ||
| b | b | a | a | b | b | b | a | c | b | b | a | b | a | B | b | a | a | b | ||
| a | b | a | b | b | b | b | a | c | b | d | a | NR | b | - | d | c | c | b | ||
| a | a | a | c | e | b | b | b | c | b | b | a | a | b | - | b | b | a | a | ||
| d | b | a | a | b | a | b | b | b | b | b | b | b | a | A | c | c | c | b | ||
| b | a | a | c | d | b | b | a | b | b | c | a | c | b | C | c | a | a | b | ||
| c | b | a | a | b | b | b | a | NA | NA | NA | NA | b | NA | NA | c | NA | NA | NA | ||
| MBL- medical biochemistry laboratory. Q1 - Q18 – questions defined in | ||||||||||||||||||||
Figure 1Electrodes for pilocarpine iontophoresis (PI) produced by local manufacturer