| Literature DB >> 28390480 |
Shotaro Takano1, Masaharu Tanimizu2, Takafumi Hirata3, Ki-Cheol Shin4, Yusuke Fukami5, Katsuhiko Suzuki5, Yoshiki Sohrin6.
Abstract
Stable isotope ratios of nickel, copper, and zinc are powerful tools for elucidating the biogeochemical cycling of trace metals in the ocean. However, analytical difficulties have impeded isotopic studies of these metals. We present a simple and rapid method for simultaneous analysis of Ni, Cu, and Zn isotope ratios in seawater using NOBIAS Chelate-PA1 resin and anion exchange resin. A NOBIAS Chelate-PA1 resin column was used to quantitatively collect Ni, Cu, and Zn from seawater and thoroughly remove the seawater matrix. Subsequent anion exchange purified and separated the Ni, Cu, and Zn from each other. The blanks used in this method (0.22 ng for Ni, 0.29 ng for Cu, and 0.53 ng for Zn) were sufficiently low to determine the isotope ratios of Ni, Cu, and Zn in surface seawater. Using this method, we analyzed GEOTRACES reference seawater samples (i.e., SAFe D1 and SAFe D2), National Research Council Canada certified materials (i.e., CASS-5 and NASS-6), and seawater samples collected from different depths in the subarctic South Pacific. The results were consistent with previously reported values. This method is expected to accelerate isotopic research and contribute to our understanding of biogeochemical cycling in the ocean.Entities:
Keywords: Anion exchange; Isotope ratio; NOBIAS; Ocean; Reference material; Trace metals
Year: 2017 PMID: 28390480 DOI: 10.1016/j.aca.2017.03.010
Source DB: PubMed Journal: Anal Chim Acta ISSN: 0003-2670 Impact factor: 6.558