Biomechanical Assessment of the Dorsal Spanning Bridge Plate in Distal Radius Fracture Fixation: Implications for Immediate Weight-Bearing.

Jerry I Huang1, Bret Peterson1, Kate Bellevue1, Nicolas Lee2, Sean Smith2, Safa Herfat2.   

Abstract

BACKGROUND: The goal of this study was to compare the biomechanical stability of a 2.4-mm dorsal spanning bridge plate with a volar locking plate (VLP) in a distal radius fracture model, during simulated crutch weight-bearing.
METHODS: Five paired cadaveric forearms were tested. A 1-cm dorsal wedge osteotomy was created to simulate an unstable distal radius fracture with dorsal comminution. Fractures were fixed with a VLP or a dorsal bridge plate (DBP). Specimens were mounted to a crutch handle, and optical motion-tracking sensors were attached to the proximal and distal segments. Specimens were loaded in compression at 1 mm/s on a servohydraulic test frame until failure, defined as 2 mm of gap site displacement.
RESULTS: The VLP construct was significantly more stable to axial load in a crutch weight-bearing model compared with the DBP plate (VLP: 493 N vs DBP: 332 N). Stiffness was higher in the VLP constructs, but this was not statistically significant (VLP: 51.4 N/mm vs DBP: 32.4 N/mm). With the crutch weight-bearing model, DBP failed consistently with wrist flexion and plate bending, whereas VLP failed with axial compression at the fracture site and dorsal collapse.
CONCLUSIONS: Dorsal spanning bridge plating is effective as an internal spanning fixator in treating highly comminuted intra-articular distal radius fracture and prevents axial collapse at the radiocarpal joint. However, bridge plating may not offer advantages in early weight-bearing or transfer in polytrauma patients, with less axial stability in our crutch weight-bearing model compared with volar plating. A stiffer 3.5-mm DBP or use of a DBP construct without the central holes may be considered for distal radius fractures if the goal is early crutch weight-bearing through the injured extremity.

Entities:  

Keywords:  biomechanical study; distal radius fracture; distraction plating; dorsal bridge plating; early weight-bearing

Mesh:

Year:  2017        PMID: 28387161      PMCID: PMC5987984          DOI: 10.1177/1558944717701235

Source DB:  PubMed          Journal:  Hand (N Y)        ISSN: 1558-9447


  13 in total

1.  Complications associated with distraction plate fixation of wrist fractures.

Authors:  Douglas P Hanel; Scott David Ruhlman; Leo I Katolik; Christopher H Allan
Journal:  Hand Clin       Date:  2010-05       Impact factor: 1.907

2.  The effects of screw length on stability of simulated osteoporotic distal radius fractures fixed with volar locking plates.

Authors:  Lindley B Wall; Michael D Brodt; Matthew J Silva; Martin I Boyer; Ryan P Calfee
Journal:  J Hand Surg Am       Date:  2012-02-02       Impact factor: 2.230

Review 3.  Epidemiology of osteoporosis and osteoporotic fractures.

Authors:  S R Cummings; J L Kelsey; M C Nevitt; K J O'Dowd
Journal:  Epidemiol Rev       Date:  1985       Impact factor: 6.222

4.  A biomechanic comparison of an internal radiocarpal-spanning 2.4-mm locking plate and external fixation in a model of distal radius fractures.

Authors:  Jonathon C Wolf; Wayne M Weil; Douglas P Hanel; Thomas E Trumble
Journal:  J Hand Surg Am       Date:  2006-12       Impact factor: 2.230

5.  Bridge plating of distal radius fractures: the Harborview method.

Authors:  Douglas P Hanel; Thomas S Lu; Wayne M Weil
Journal:  Clin Orthop Relat Res       Date:  2006-04       Impact factor: 4.176

6.  Use of a distraction plate for distal radial fractures with metaphyseal and diaphyseal comminution.

Authors:  David S Ruch; T Adam Ginn; Charles C Yang; Beth P Smith; Julia Rushing; Douglas P Hanel
Journal:  J Bone Joint Surg Am       Date:  2005-05       Impact factor: 5.284

7.  Distraction plating for the treatment of highly comminuted distal radius fractures in elderly patients.

Authors:  Marc J Richard; Leonid I Katolik; Douglas P Hanel; Daniel A Wartinbee; David S Ruch
Journal:  J Hand Surg Am       Date:  2012-04-04       Impact factor: 2.230

8.  Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models.

Authors:  Justin M Uhl; Bernard Seguin; Amy S Kapatkin; Kurt S Schulz; Tanya C Garcia; Susan M Stover
Journal:  Vet Surg       Date:  2008-10       Impact factor: 1.495

Review 9.  Use of a distraction plate for distal radial fractures with metaphyseal and diaphyseal comminution. Surgical technique.

Authors:  T Adam Ginn; David S Ruch; Charles C Yang; Douglas P Hanel
Journal:  J Bone Joint Surg Am       Date:  2006-03       Impact factor: 5.284

10.  Biomechanical comparison of osteoporotic distal radius fractures fixed by distal locking screws with different length.

Authors:  Xiong Liu; Wei-dong Wu; Ya-feng Fang; Mei-chao Zhang; Wen-hua Huang
Journal:  PLoS One       Date:  2014-07-31       Impact factor: 3.240

View more
  2 in total

1.  Bridge Plate Design Effects on Yield and Fatigue in Distal Radius Fracture Model.

Authors:  Asif M Ilyas; Gerald M Hayward; Jonathan A Harris; Wenhai Wang; Brandon S Bucklen
Journal:  J Wrist Surg       Date:  2020-07-06

Review 2.  Review of Internal Radiocarpal Distraction Plating for Distal Radius Fracture Fixation.

Authors:  Venus Vakhshori; Ram Kiran Alluri; Milan Stevanovic; Alidad Ghiassi
Journal:  Hand (N Y)       Date:  2018-07-13