Evaluation of four hemoglobin separation analyzers for hemoglobinopathy diagnosis.

Simon Degandt1, Rémy Coens1, Barbara Cauwelier1, Helena Devos1, Michel Langlois1, Jan Emmerechts1.   

Abstract

BACKGROUND: Four automated hemoglobin separation devices are compared in their ability to detect hemoglobinopathies, both in HbA1c and in hemoglobinopathy mode.
METHODS: Quality control material and 58 samples, including one heterozygous α-thalassemia sample, six heterozygote β-thalassemia samples and 32 samples with a known hemoglobin variant, were used to assess imprecision of HbF and HbA2 measurements, correlation with the gold standard and sensitivity for detecting β-thalassemia and Hb variants on D-100 (Bio-Rad Laboratories), HA 8180T (Menarini), HLC-723G8 (Tosoh Bioscience) and Capillarys 2 Flex Piercing (Sebia).
RESULTS: Imprecision was <10% for both HbF and HbA2 in all modes of all analyzers. Correlation studies for HbF and HbA2 demonstrated statistically significant but small biases when compared to the gold standard. All six β-thalassemia samples but one were detected on all analyzers using a HbA2 cut-off value of 3.5%. D-100, HA8180T and the Hb-pathy mode of the HLC-723G8 and the Capillarys are able to detect the most common important Hb variants (Hb C, D, E and S), but more seldom variants can be missed as they co-elute with HbA0. The HbA1c mode of the Capillarys correctly detected all measured hemoglobin variants and can therefore be used as a hemoglobinopathy screening device. This was also the case for the most common important Hb variants on the HbA1c mode of the HLC-723G8, but two rare variants were not detected.
CONCLUSION: This study stresses the importance for individual laboratories to know the advantages and drawbacks of their hemoglobin separation analyzer and its different modes in the diagnosis of hemoglobinopathies.
© 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  Bio-Rad Laboratories D-100; Menarini HA 8180T; Sebia Capillarys 2 Flex Piercing; Tosoh Bioscience HLC-723G8; hemoglobinopathy screening

Mesh:

Substances:

Year:  2017        PMID: 28383138      PMCID: PMC6817248          DOI: 10.1002/jcla.22224

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  8 in total

1.  Comparison of Sebia Capillarys Flex capillary electrophoresis with the BioRad Variant II high pressure liquid chromatography in the evaluation of hemoglobinopathies.

Authors:  Dina N Greene; Amy L Pyle; Judy S Chang; Carolyn Hoke; Thomas Lorey
Journal:  Clin Chim Acta       Date:  2012-04-10       Impact factor: 3.786

2.  Evaluating five dedicated automatic devices for haemoglobinopathy diagnostics in multi-ethnic populations.

Authors:  P Van Delft; E Lenters; M Bakker-Verweij; M de Korte; U Baylan; C L Harteveld; P C Giordano
Journal:  Int J Lab Hematol       Date:  2009-04-17       Impact factor: 2.877

3.  Evaluation of an automated capillary electrophoresis system in the screening for hemoglobinopathies.

Authors:  Frederic Cotton; Xavier Malaviolle; Françoise Vertongen; Béatrice Gulbis
Journal:  Clin Lab       Date:  2009       Impact factor: 1.138

Review 4.  ICSH recommendations for assessing automated high-performance liquid chromatography and capillary electrophoresis equipment for the quantitation of HbA2.

Authors:  A D Stephens; R Colah; S Fucharoen; J Hoyer; D Keren; A McFarlane; D Perrett; B J Wild
Journal:  Int J Lab Hematol       Date:  2015-10       Impact factor: 2.877

5.  Multicenter validation of fully automated capillary electrophoresis method for diagnosis of thalassemias and hemoglobinopathies in Thailand.

Authors:  Somchai Sangkitporn; Siripakorn K Sangkitporn; Sansanee Tanjatham; Boonnipa Suwannakan; Suthatip Rithapirom; Chonlada Yodtup; Amara Yowang; Sawitree Duangruang
Journal:  Southeast Asian J Trop Med Public Health       Date:  2011-09       Impact factor: 0.267

6.  Inherited haemoglobin disorders: an increasing global health problem.

Authors:  D J Weatherall; J B Clegg
Journal:  Bull World Health Organ       Date:  2001-10-24       Impact factor: 9.408

Review 7.  Laboratory investigation of hemoglobinopathies and thalassemias: review and update.

Authors:  G M Clarke; T N Higgins
Journal:  Clin Chem       Date:  2000-08       Impact factor: 8.327

8.  Evaluation of the Sebia Minicap Flex Piercing capillary electrophoresis for hemoglobinopathy testing.

Authors:  M Oyaert; C Van Laer; H Claerhout; P Vermeersch; K Desmet; S Pauwels; D Kieffer
Journal:  Int J Lab Hematol       Date:  2014-10-16       Impact factor: 2.877

  8 in total
  1 in total

1.  Evaluation on the separated effect of 13 hemoglobin variants by a new automatic Hba1c analyzer.

Authors:  Rui Li; Huamei Tang; Lijuan Kan; Dan Xiong; Xuedi Cheng; Yan Xu; Xiancheng Liu; Xiuming Zhang
Journal:  J Clin Lab Anal       Date:  2020-07-12       Impact factor: 2.352

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.