Literature DB >> 28381739

A Comparison between the Instantaneous Wave-free Ratio and Resting Distal Coronary Artery Pressure/Aortic Pressure and the Fractional Flow Reserve: The Diagnostic Accuracy Can Be Improved by the Use of both Indices.

Nobuo Shiode1, Tomokazu Okimoto, Hiromichi Tamekiyo, Tomoharu Kawase, Kenichi Yamane, Yuzo Kagawa, Yuto Fujii, Yusuke Ueda, Naoya Hironobe, Yasuko Kato, Yasuhiko Hayashi.   

Abstract

Objectives The fractional flow reserve (FFR) is an index of the severity of coronary stenosis that has been clinically validated in several studies. The instantaneous wave-free ratio (iFR) and the resting distal coronary artery pressure/aortic pressure (Pd/Pa) are nonhyperemic pressure-derived indices of the severity of stenosis. This study sought to examine the diagnostic accuracy of the iFR and resting Pd/Pa with respect to hyperemic FFR. Methods Following an intracoronary injection of papaverine, the iFR, resting Pd/Pa, and FFR were continuously measured in 123 lesions in 103 patients with stable coronary disease. Results The iFR and resting Pd/Pa values were strongly correlated with the FFR (R=0.794, p<0.001, R=0.832, p<0.0001, respectively). A receiver operator curve (ROC) analysis revealed that the optimal iFR cut-off value for predicting an FFR of <0.80 was 0.89 (AUC 0.901, sensitivity 84.1%, specificity 80.0%, positive predictive value 69.8%, negative predictive value 90.0%, diagnostic accuracy 81.3%), while the optimal resting Pd/Pa cut-off value was 0.92 (AUC 0.925, sensitivity 90.9%, specificity 78.5%, positive predictive value 70.2%, negative predictive value 93.9%, diagnostic accuracy 82.9%). The lesions with an iFR value of ≤0.89 and a Pd/Pa value of ≤0.92 were defined as double-positive lesions, while the lesions with an iFR value of >0.89 and a Pd/Pa value of >0.92 were defined as double-negative lesions. In these 109 lesions, the sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy were 92.3%, 82.9%, 75.0%, 95.1%, and 86.2%, respectively. Conclusion This analysis demonstrated that the iFR and resting Pd/Pa were strongly correlated with the FFR and that the diagnostic accuracy of the iFR was similar to that of the resting Pd/Pa. The diagnostic accuracy can be improved with the use of both the iFR and the resting Pd/Pa.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28381739      PMCID: PMC5457916          DOI: 10.2169/internalmedicine.56.7857

Source DB:  PubMed          Journal:  Intern Med        ISSN: 0918-2918            Impact factor:   1.271


Introduction

The fundamental limitations of coronary angiography and its poor correlation with the severity of functional stenosis, in terms of blood flow, are well recognized (1). The fractional flow reserve (FFR) is an accurate invasive index that can be used in a catheterization laboratory to determine whether angiographically-equivocal stenosis is of functional significance (2). Recent randomized trials assessing the value of FFR-guided PCI have demonstrated reduced rates of major adverse cardiac events, mainly due to a decreased need for repeat revascularization (3, 4). These results have led to changes in practice guidelines. The use of the FFR is recommended for the assessment of lesions of intermediate severity (5, 6). With regard to the measurement of the FFR, some drugs are currently used for producing maximal coronary hyperemia in humans. The bolus intracoronary administration of papaverine or the intravenous infusion of adenosine or adenosine triphosphate (ATP) is capable of producing maximal coronary hyperemia (7-9). However, it takes much more time and cost to induce pharmacological hyperemia and some patients experience chest discomfort during hyperemia. Two nonhyperemic measures of inducing pressure might be useful for assessing the severity of coronary stenosis. The resting distal coronary artery pressure/aortic pressure (Pd/Pa) is the ratio of distal coronary artery pressure to aortic pressure over the entire cardiac cycle. Conversely, the instantaneous wave-free ratio (iFR) measures coronary pressure during a specific period of diastole when the resting resistance is the lowest (10). The assessment of the severity of coronary stenosis without the induction of hyperemia is attractive because it reducing the procedural time and cost, and avoids the patient-related discomfort associated with pharmacological hyperemia. This study sought to examine the diagnostic accuracy of the iFR and resting Pd/Pa with in comparison to hyperemic FFR.

Material and Methods

Patients

One hundred three patients with mild or moderate coronary stenosis who were undergoing coronary angiography were selected for this study. The patients' characteristics are shown in Table 1. Written informed consent was obtained from all of the patients prior to coronary angiography and this study protocol was approved by the ethical committee of Tsuchiya General Hospital.
Table 1.

Patients Characteristics.

n=103
age (years)70.4±8.7
Male (%)77 (74.8%)
Cardiac Status
Stable Angina92
Previous MI11
Risk Factors
Hypertension82
Dyslipidemia61
Diabetes40
Smoking28
Hemodialysis12

MI: myocardial infarction

Patients Characteristics. MI: myocardial infarction

Study design

Diagnostic coronary angiography was performed through a standard percutaneous radial or femoral arterial approach. After obtaining vascular access, 3,000 units of heparin were administered intravenously. A 6F guide catheter was introduced into the left or right coronary arteries. The FFR was measured with a coronary pressure guide wire (Verrata guide wire; Volcano Corporation, San Diego, CA) as follows: 1) resting Pd/Pa; 2) iFR; 3) FFR after the intracoronary injection of papaverine (12 mg in the left coronary artery or 8 mg in the right coronary artery). The FFR, electrocardiography results and arterial blood pressure were monitored until the FFR value returned to the baseline value.

Quantitative coronary angiography analysis (QCA)

The results from the single most severe view were recorded. The lesion length, minimum lumen diameter (MLD), reference vessel diameter (RVD) and percent diameter stenosis (%DS) were analyzed using a computerized, automated, edge detection algorithm (Philips Medical System, Best, The Netherlands), as previously described (11).

Statistical analysis

Continuous variables were expressed as the mean±SD and compared using the paired t-test. The performance of the iFR and the resting Pd/Pa in predicting a positive FFR (<0.80) was assessed in terms of the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and diagnostic accuracy, together with their 95% confidence interval (CI). A receiver operating characteristic (ROC) curve was created to measure the accuracy of the test (area under the ROC curve: AUC) and identify iFR and the optimal resting Pd/Pa cut-off value that corresponded to an FFR of <0.80.

Results

The characteristics of the patients and lesions

The patients' characteristics are summarized in Table 1. Eleven patients had a past history of myocardial infarction. Twelve patients had end-stage renal dysfunction with hemodialysis. The characteristics of the lesion are summarized in Table 2. The locations of the lesions were as follows: left anterior descending coronary artery (LAD) (n=90), left circumflex coronary artery (LCx) (n=4), and right coronary artery (RCA) (n=29). The mean % stenosis was 49.1±16.1%.
Table 2.

Lesions’ Characteritics.

Lesion Locationn=123
LAD90
LCx4
RCA29
QCA data
Lesion Length (mm)9.6±5.4
RVD (mm)2.67±0.65
MLD (mm)1.36±0.50
% stenosis49.1±16.1

LAD: left anterior descending artery

LCx: left circumflex artery, RCA: right coronary artery

RVD: reference vessel diameter

MLD: minimum lesion diameter

Lesions’ Characteritics. LAD: left anterior descending artery LCx: left circumflex artery, RCA: right coronary artery RVD: reference vessel diameter MLD: minimum lesion diameter

The correlations between the iFR and FFR values and between the resting Pd/Pa and FFR values

Both the iFR and resting Pd/Pa were strongly correlated with the FFR. The correlation between the iFR and the FFR was R=0.794 (p<0.0001, FFR=0.12+0.77iFR). The correlation between the resting Pd/Pa and the FFR was R=0.832 (p<0.0001, FFR=-0.34+1.25Pd/Pa) (Fig. 1). The receiver operator curve (ROC) analysis revealed that the optimal iFR cut-off value for predicting an FFR value of <0.80 was 0.89 (AUC 0.901, sensitivity 84.1%, specificity 80.0%, positive predictive value 69.8%, negative predictive value 90.0%, diagnostic accuracy 81.3%), while the optimal resting Pd/Pa cut-off value was 0.92 (AUC 0.925, sensitivity 90.9%, specificity 78.5%, positive predictive value 70.2%, negative predictive value 93.9%, diagnostic accuracy 82.9%) (Fig. 2). The lesions with an iFR value of ≤0.89 and a Pd/Pa value of ≤0.92 were defined as double-positive, while those with an iFR value of >0.89 and a Pd/Pa value of >0.92 were defined as double-negative lesions. These lesions were summarized in Table 3. In these 109 lesions, the sensitivity, specificity, positive predictive value, negative predictive value and diagnostic accuracy were 92.3%, 82.9%, 75.0%, 95.1%, 86.2%, respectively.
Figure 1.

The correlations between the iFR, Pd/Pa and FFR.

Figure 2.

The results of the receiver operator curve (ROC) analysis. The optimal iFR and Pd/Pa cut-off values for predicting an FFR of <0.80 were 0.89 (AUC 0.901) and 0.92 (AUC 0.925), respectively.

Table 3.

Double Positive and Double Negative Lesions.

FFR<0.80FFR≥0.80
iFR≤0.89 and Pd/Pa≤0.923612
iFR>0.89 and Pd/Pa>0.92356
The correlations between the iFR, Pd/Pa and FFR. The results of the receiver operator curve (ROC) analysis. The optimal iFR and Pd/Pa cut-off values for predicting an FFR of <0.80 were 0.89 (AUC 0.901) and 0.92 (AUC 0.925), respectively. Double Positive and Double Negative Lesions.

Deviated lesions

The lesions in which the actual measured iFR and resting Pd/Pa values differed from the predicted iFR and resting Pd/Pa values by >10% (as calculated using the formula: FFR=0.12+0.77 iFR, FFR=-0.34+1.25 Pd/Pa) were defined as deviated lesions. The lesions in which the difference between the actual measured iFR and the resting Pd/Pa values and the predicted values were <10% were defined as non-deviated lesions. The lesions in which the actual measured iFR and the resting Pd/Pa was >10% higher (iFR, n=11; resting Pd/Pa, n=11) or >10% lower (iFR, n=6; resting Pd/Pa, n=4) than the predicted iFR and resting Pd/Pa values were defined as underestimated lesions and overestimated lesions, respectively. We compared the QCA data of the deviated lesions with those of the non-deviated lesions. In the lesions underestimated by iFR, the lesion length was significantly longer and both the RVD and MLD were significantly smaller in comparison to the non-deviated lesions (Table 4). In the lesions that were underestimated by the resting Pd/Pa, the lesion length was significantly longer, both the RVD and MLD were significantly smaller and the % stenosis was more severe in comparison to the non-deviated lesions (Table 5). Thus, in small vessels with long, severe stenosis, both the iFR and the resting Pd/Pa tended to be larger than the values predicted from the FFR.
Table 4.

Deviated Lesions in iFR.

Underestimated LesionsOverestimated LesionsNon-Deviated Lesions
n116106
Lesion Length (mm)12.9±9.0*8.3±5.99.3±4.8
RVD (mm)2.32±0.59*2.37±0.422.73±0.65
MLD (mm)0.96±0.45**1.31±0.441.40±0.65
%stenosis58.1±19.744.1±17.648.5±15.5

RVD: reference vessel diameter, MLD: minimum lesion diameter

*p<0.05, **p<0.01 vs. Corrected Estimated Lesions

Table 5.

Deviated Lesions in Resting Pd/Pa.

Underestimated LesionsOverestimated LesionsNon-Deviated Lesions
n114108
Lesion Length (mm)12.6±8.5*11.1±5.39.3±4.9
RVD (mm)2.40±0.532.48±0.452.71±0.66
MLD (mm)0.94±0.35**1.20±0.171.40±0.50
%stenosis61.0±13.5*50.7±9.647.8±16.1

RVD: reference vessel diameter, MLD: minimum lesion diameter

*p<0.05, **p<0.01 vs. Corrected Estimated Lesions

Deviated Lesions in iFR. RVD: reference vessel diameter, MLD: minimum lesion diameter *p<0.05, **p<0.01 vs. Corrected Estimated Lesions Deviated Lesions in Resting Pd/Pa. RVD: reference vessel diameter, MLD: minimum lesion diameter *p<0.05, **p<0.01 vs. Corrected Estimated Lesions

Hemodynamics

The patients' systolic and diastolic blood pressure values significantly decreased after the injection of papaverine, while their heart rates significantly increased (Table 6).
Table 6.

Hemodynamic Change.

iFRPd/Papapaverine
sys BP (mmHg)130±21131±22120±19*
dia BP (mmHg)67±1067±1061±9*
HR (beats/min)70±1171±1175±12*

sys BP: systolic blood pressure, dia BP: diastolic blood pressure

HR: heart rate, *p<0.01 vs. iFR and Pd/Pa

Hemodynamic Change. sys BP: systolic blood pressure, dia BP: diastolic blood pressure HR: heart rate, *p<0.01 vs. iFR and Pd/Pa

Discussion

Two nonhyperemic measures of pressure might be useful for assessing the severity of coronary stenosis. Pd/Pa is the ratio of distal coronary artery pressure to aortic pressure over the entire cardiac cycle. Conversely, the iFR measures coronary pressure during a specific period of diastole when resting resistance is the lowest. In this study the overall linear correlation between both iFR and resting Pd/Pa and FFR was moderate, and both of the nonhyperemic indices showed moderate overall diagnostic accuracy. However, in prior reports, the diagnostic accuracy of the iFR (in comparison to the FFR) ranged widely, from 60% to 91% (10, 12-14), and its relative accuracy in comparison to the resting Pd/Pa has been debated. In the present study, in small vessels with severe, long stenosis both the iFR and the resting Pd/Pa values tended to be higher than the predicted value that was calculated from the actual measured FFR value. In other words, the degree of functional stenosis tended to be underestimated in small vessels with severe, long stenosis. It would be better to carefully evaluate small vessels with long and severe stenosis using both iFR and resting Pd/Pa. The diagnostic accuracy of iFR alone or resting Pd/Pa alone was moderate. However, if the lesions with an iFR ≤0.89 and a resting Pd/Pa of ≤0.92 were defined as double-positive and those with an iFR of >0.89 and a resting Pd/Pa of >0.92 were defined as double-negative, the diagnostic accuracy improved. Although the iFR and the resting Pd/Pa are imperfect surrogates of the FFR when they are close to the clinically used cutoff value of 0.80, the diagnostic accuracy can be improved by using both indices together. Only 14 lesions (11.4%) were neither double-positive nor double-negative. Thus, in clinical practice, the lesions can be deferred if they are found to be double-negative by both iFR and resting Pd/Pa. Only the lesions that were neither double-positive nor double-negative should be evaluated by FFR with drug-induced hyperemia. Several investigators have reported the relationship between conventional FFR and the non-hyperemic indices of the severity of stenosis (13-15). In the RESOLVE study, the overall linear correlation between both the iFR and the Pd/Pa and the FFR was strong (R value 0.81 and 0.83, respectively), and the optimal cut points for predicting an FFR of <0.80 were 0.90 and 0.92 respectively. The diagnostic accuracy was 80.4% and 81.5% respectively. These results were similar to those of our own study. Based on the level of accuracy, it seems to be insufficient to separately use either of the parameters for procedural guidance in all cases because approximately 20% of the therapeutic decisions would be discordant from the FFR. However, we are of the opinion that the combined use both the iFR and the resting Pd/Pa may have the potential to improve the diagnostic accuracy in clinical practice. The present study is associated with several limitations. Although the data were collected prospectively with the application of a strict study methodology in patients with intermediate stenosis in the clinical setting, the present study was performed in a single center study and involved a relatively small number of patients. The analyses of the angiograms, ECGs, and pressure waveforms were performed at an in-hospital lab in a blinded fashion, but not by a core laboratory. The RESOLVE study is the first coronary physiology study to have used a core laboratory for the analysis of the diagnostic accuracy of hyperemic and resting pressure-derived indices of the severity of stenosis. In the RESOLVE study, 19.3% of measurements were found to be suboptimal and were excluded from the analysis (15). Future clinical studies should consider including a core laboratory analysis.

Conclusion

Both the iFR and resting Pd/Pa were strongly correlated with the FFR and the diagnostic accuracy of the iFR (81.3%) was similar to that of the resting Pd/Pa (82.9%). In small vessels with severe, long stenosis the degree of functional stenosis tended to be underestimated by both indices. The diagnostic accuracy could be improved by the combined use of both the iFR and the resting Pd/Pa.

The authors state that they have no Conflict of Interest (COI).
  15 in total

1.  Development and validation of a new adenosine-independent index of stenosis severity from coronary wave-intensity analysis: results of the ADVISE (ADenosine Vasodilator Independent Stenosis Evaluation) study.

Authors:  Sayan Sen; Javier Escaned; Iqbal S Malik; Ghada W Mikhail; Rodney A Foale; Rafael Mila; Jason Tarkin; Ricardo Petraco; Christopher Broyd; Richard Jabbour; Amarjit Sethi; Christopher S Baker; Micheal Bellamy; Mahmud Al-Bustami; David Hackett; Masood Khan; David Lefroy; Kim H Parker; Alun D Hughes; Darrel P Francis; Carlo Di Mario; Jamil Mayet; Justin E Davies
Journal:  J Am Coll Cardiol       Date:  2011-12-07       Impact factor: 24.094

2.  2013 ESC guidelines on the management of stable coronary artery disease: the Task Force on the management of stable coronary artery disease of the European Society of Cardiology.

Authors:  Gilles Montalescot; Udo Sechtem; Stephan Achenbach; Felicita Andreotti; Chris Arden; Andrzej Budaj; Raffaele Bugiardini; Filippo Crea; Thomas Cuisset; Carlo Di Mario; J Rafael Ferreira; Bernard J Gersh; Anselm K Gitt; Jean-Sebastien Hulot; Nikolaus Marx; Lionel H Opie; Matthias Pfisterer; Eva Prescott; Frank Ruschitzka; Manel Sabaté; Roxy Senior; David Paul Taggart; Ernst E van der Wall; Christiaan J M Vrints; Jose Luis Zamorano; Stephan Achenbach; Helmut Baumgartner; Jeroen J Bax; Héctor Bueno; Veronica Dean; Christi Deaton; Cetin Erol; Robert Fagard; Roberto Ferrari; David Hasdai; Arno W Hoes; Paulus Kirchhof; Juhani Knuuti; Philippe Kolh; Patrizio Lancellotti; Ales Linhart; Petros Nihoyannopoulos; Massimo F Piepoli; Piotr Ponikowski; Per Anton Sirnes; Juan Luis Tamargo; Michal Tendera; Adam Torbicki; William Wijns; Stephan Windecker; Juhani Knuuti; Marco Valgimigli; Héctor Bueno; Marc J Claeys; Norbert Donner-Banzhoff; Cetin Erol; Herbert Frank; Christian Funck-Brentano; Oliver Gaemperli; José R Gonzalez-Juanatey; Michalis Hamilos; David Hasdai; Steen Husted; Stefan K James; Kari Kervinen; Philippe Kolh; Steen Dalby Kristensen; Patrizio Lancellotti; Aldo Pietro Maggioni; Massimo F Piepoli; Axel R Pries; Francesco Romeo; Lars Rydén; Maarten L Simoons; Per Anton Sirnes; Ph Gabriel Steg; Adam Timmis; William Wijns; Stephan Windecker; Aylin Yildirir; Jose Luis Zamorano
Journal:  Eur Heart J       Date:  2013-08-30       Impact factor: 29.983

3.  Serious ventricular dysrhythmias after intracoronary papaverine.

Authors:  R F Wilson; C W White
Journal:  Am J Cardiol       Date:  1988-12-01       Impact factor: 2.778

4.  ACCF/SCAI/STS/AATS/AHA/ASNC/HFSA/SCCT 2012 Appropriate use criteria for coronary revascularization focused update: a report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, Society for Cardiovascular Angiography and Interventions, Society of Thoracic Surgeons, American Association for Thoracic Surgery, American Heart Association, American Society of Nuclear Cardiology, and the Society of Cardiovascular Computed Tomography.

Authors:  Manesh R Patel; Gregory J Dehmer; John W Hirshfeld; Peter K Smith; John A Spertus
Journal:  J Am Coll Cardiol       Date:  2012-01-30       Impact factor: 24.094

5.  Does visual interpretation of the coronary arteriogram predict the physiologic importance of a coronary stenosis?

Authors:  C W White; C B Wright; D B Doty; L F Hiratza; C L Eastham; D G Harrison; M L Marcus
Journal:  N Engl J Med       Date:  1984-03-29       Impact factor: 91.245

6.  Effect of adenosine triphosphate on human coronary circulation.

Authors:  N Shiode; M Kato; K Nakayama; K Shinohara; J Kurokawa; T Yamagata; H Matsuura; G Kajiyama
Journal:  Intern Med       Date:  1998-10       Impact factor: 1.271

7.  Does the instantaneous wave-free ratio approximate the fractional flow reserve?

Authors:  Nils P Johnson; Richard L Kirkeeide; Kaleab N Asrress; William F Fearon; Timothy Lockie; Koen M J Marques; Stylianos A Pyxaras; M Cristina Rolandi; Marcel van 't Veer; Bernard De Bruyne; Jan J Piek; Nico H J Pijls; Simon Redwood; Maria Siebes; Jos A E Spaan; K Lance Gould
Journal:  J Am Coll Cardiol       Date:  2013-02-06       Impact factor: 24.094

8.  Classification performance of instantaneous wave-free ratio (iFR) and fractional flow reserve in a clinical population of intermediate coronary stenoses: results of the ADVISE registry.

Authors:  Ricardo Petraco; Javier Escaned; Sayan Sen; Sukhjinder Nijjer; Kaleab N Asrress; Mauro Echavarria-Pinto; Tim Lockie; Muhammed Z Khawaja; Cecilia Cuevas; Nicolas Foin; Christopher Broyd; Rodney A Foale; Nearchos Hadjiloizou; Iqbal S Malik; Ghada W Mikhail; Amarjit Sethi; Raffi Kaprielian; Christopher S Baker; David Lefroy; Michael Bellamy; Mahmud Al-Bustami; Masood A Khan; Alun D Hughes; Darrel P Francis; Jamil Mayet; Carlo Di Mario; Simon Redwood; Justin E Davies
Journal:  EuroIntervention       Date:  2013-05-20       Impact factor: 6.534

9.  Multicenter core laboratory comparison of the instantaneous wave-free ratio and resting Pd/Pa with fractional flow reserve: the RESOLVE study.

Authors:  Allen Jeremias; Akiko Maehara; Philippe Généreux; Kaleab N Asrress; Colin Berry; Bernard De Bruyne; Justin E Davies; Javier Escaned; William F Fearon; K Lance Gould; Nils P Johnson; Ajay J Kirtane; Bon-Kwon Koo; Koen M Marques; Sukhjinder Nijjer; Keith G Oldroyd; Ricardo Petraco; Jan J Piek; Nico H Pijls; Simon Redwood; Maria Siebes; Jos A E Spaan; Marcel van 't Veer; Gary S Mintz; Gregg W Stone
Journal:  J Am Coll Cardiol       Date:  2013-11-06       Impact factor: 24.094

10.  VERIFY (VERification of Instantaneous Wave-Free Ratio and Fractional Flow Reserve for the Assessment of Coronary Artery Stenosis Severity in EverydaY Practice): a multicenter study in consecutive patients.

Authors:  Colin Berry; Marcel van 't Veer; Nils Witt; Petr Kala; Otakar Bocek; Stylianos A Pyxaras; John D McClure; William F Fearon; Emanuele Barbato; Pim A L Tonino; Bernard De Bruyne; Nico H J Pijls; Keith G Oldroyd
Journal:  J Am Coll Cardiol       Date:  2013-02-06       Impact factor: 24.094

View more
  5 in total

1.  The Importance of Measuring Coronary Blood Flow for Clinical Decision Making.

Authors:  Cynthia Yeung; Adrian Baranchuk; Gary Tse; Tong Liu
Journal:  Curr Cardiol Rev       Date:  2019

2.  Meta-Analysis of Diagnostic Performance of Instantaneous Wave-Free Ratio versus Quantitative Flow Ratio for Detecting the Functional Significance of Coronary Stenosis.

Authors:  Wenjie Zuo; Mingming Yang; Yifan Chen; Aiming Xie; Lijuan Chen; Genshan Ma
Journal:  Biomed Res Int       Date:  2019-04-18       Impact factor: 3.411

3.  Effect of Elevated Left Ventricular End Diastolic Pressure on Instantaneous Wave-Free Ratio and Fractional Flow Reserve Discordance.

Authors:  Hassan Tahir; James Livesay; Benjamin Fogelson; Raj Baljepally
Journal:  Cardiol Res       Date:  2021-02-23

4.  Qualitative assessments of myocardial ischemia by cardiac MRI and coronary stenosis by invasive coronary angiography in relation to quantitative perfusion by positron emission tomography in patients with known or suspected stable coronary artery disease.

Authors:  Shahnaz Akil; Fredrik Hedeer; Marcus Carlsson; Håkan Arheden; Jenny Oddstig; Cecilia Hindorf; Jonas Jögi; David Erlinge; Henrik Engblom
Journal:  J Nucl Cardiol       Date:  2018-12-10       Impact factor: 5.952

5.  Appropriate coronary revascularization can be accomplished if myocardial perfusion is quantified by positron emission tomography prior to treatment decision.

Authors:  Shahnaz Akil; Fredrik Hedeer; Jenny Oddstig; Thomas Olsson; Jonas Jögi; David Erlinge; Marcus Carlsson; Håkan Arheden; Cecilia Hindorf; Henrik Engblom
Journal:  J Nucl Cardiol       Date:  2019-11-08       Impact factor: 5.952

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.