The impact of smart metal artefact reduction algorithm for use in radiotherapy treatment planning.

Connor Guilfoile1,2, Peter Rampant3, Michael House4.   

Abstract

The presence of metal artefacts in computed tomography (CT) create issues in radiation oncology. The loss of anatomical information and incorrect Hounsfield unit (HU) values produce inaccuracies in dose calculations, providing suboptimal patient treatment. Metal artefact reduction (MAR) algorithms were developed to combat these problems. This study provides a qualitative and quantitative analysis of the "Smart MAR" software (General Electric Healthcare, Chicago, IL, USA), determining its usefulness in a clinical setting. A detailed analysis was conducted using both patient and phantom data, noting any improvements in HU values and dosimetry with the GE-MAR enabled. This study indicates qualitative improvements in severity of the streak artefacts produced by metals, allowing for easier patient contouring. Furthermore, the GE-MAR managed to recover previously lost anatomical information. Additionally, phantom data showed an improvement in HU value with GE-MAR correction, producing more accurate point dose calculations in the treatment planning system. Overall, the GE-MAR is a useful tool and is suitable for clinical environments.

Entities:  

Keywords:  Computed tomography (CT); Dosimetry; Metal artefact reduction (MAR); Radiotherapy

Mesh:

Substances:

Year:  2017        PMID: 28378320     DOI: 10.1007/s13246-017-0543-5

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  1 in total

1.  Feasibility of Metal Artifact Reduction on CT Angiography for Planning Direct Surgery of Tentorial dAVF after Onyx Embolization.

Authors:  Kanji Nakagawa; Shingo Toyota; Takeshi Shimizu; Tomoaki Murakami; Takuyu Taki
Journal:  Asian J Neurosurg       Date:  2022-08-25
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.