| Literature DB >> 28378320 |
Connor Guilfoile1,2, Peter Rampant3, Michael House4.
Abstract
The presence of metal artefacts in computed tomography (CT) create issues in radiation oncology. The loss of anatomical information and incorrect Hounsfield unit (HU) values produce inaccuracies in dose calculations, providing suboptimal patient treatment. Metal artefact reduction (MAR) algorithms were developed to combat these problems. This study provides a qualitative and quantitative analysis of the "Smart MAR" software (General Electric Healthcare, Chicago, IL, USA), determining its usefulness in a clinical setting. A detailed analysis was conducted using both patient and phantom data, noting any improvements in HU values and dosimetry with the GE-MAR enabled. This study indicates qualitative improvements in severity of the streak artefacts produced by metals, allowing for easier patient contouring. Furthermore, the GE-MAR managed to recover previously lost anatomical information. Additionally, phantom data showed an improvement in HU value with GE-MAR correction, producing more accurate point dose calculations in the treatment planning system. Overall, the GE-MAR is a useful tool and is suitable for clinical environments.Entities:
Keywords: Computed tomography (CT); Dosimetry; Metal artefact reduction (MAR); Radiotherapy
Mesh:
Substances:
Year: 2017 PMID: 28378320 DOI: 10.1007/s13246-017-0543-5
Source DB: PubMed Journal: Australas Phys Eng Sci Med ISSN: 0158-9938 Impact factor: 1.430