Real-time intraoperative evaluation of implant quality and dose correction during prostate brachytherapy consistently improves target coverage using a novel image fusion and optimization program.

Michael J Zelefsky1, Gilad N Cohen2, Amandeep S Taggar2, Marisa Kollmeier2, Sean McBride2, Gig Mageras2, Marco Zaider2.   

Abstract

PURPOSE: Our purpose was to describe the process and outcome of performing postimplantation dosimetric assessment and intraoperative dose correction during prostate brachytherapy using a novel image fusion-based treatment-planning program. METHODS AND MATERIALS: Twenty-six consecutive patients underwent intraoperative real-time corrections of their dose distributions at the end of their permanent seed interstitial procedures. After intraoperatively planned seeds were implanted and while the patient remained in the lithotomy position, a cone beam computed tomography scan was obtained to assess adequacy of the prescription dose coverage. The implanted seed positions were automatically segmented from the cone-beam images, fused onto a new set of acquired ultrasound images, reimported into the planning system, and recontoured. Dose distributions were recalculated based upon actual implanted seed coordinates and recontoured ultrasound images and were reviewed. If any dose deficiencies within the prostate target were identified, additional needles and seeds were added. Once an implant was deemed acceptable, the procedure was completed, and anesthesia was reversed.
RESULTS: When the intraoperative ultrasound-based quality assurance assessment was performed after seed placement, the median volume receiving 100% of the dose (V100) was 93% (range, 74% to 98%). Before seed correction, 23% (6/26) of cases were noted to have V100 <90%. Based on this intraoperative assessment and replanning, additional seeds were placed into dose-deficient regions within the target to improve target dose distributions. Postcorrection, the median V100 was 97% (range, 93% to 99%). Following intraoperative dose corrections, all implants achieved V100 >90%.
CONCLUSIONS: In these patients, postimplantation evaluation during the actual prostate seed implant procedure was successfully applied to determine the need for additional seeds to correct dose deficiencies before anesthesia reversal. When applied, this approach should significantly reduce intraoperative errors and chances for suboptimal dose delivery during prostate brachytherapy.
Copyright © 2017 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28377139     DOI: 10.1016/j.prro.2017.01.009

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  2 in total

Review 1.  Low dose rate prostate brachytherapy.

Authors:  Bradley J Stish; Brian J Davis; Lance A Mynderse; Robert H McLaren; Christopher L Deufel; Richard Choo
Journal:  Transl Androl Urol       Date:  2018-06

2.  Dosimetric characterization of GMS BT-125-1 125 I radioactive seed with Monte Carlo simulations and experimental measurement.

Authors:  Nan Zhao; Ruijie Yang; Li Ren; Yi Fan; Junli Li; Jianguo Zhang
Journal:  J Appl Clin Med Phys       Date:  2017-09-14       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.