Literature DB >> 28376646

Why the Diabetes Technology Society Surveillance Protocol for Glucose Meters Needs to Be Revised.

Jan S Krouwer1.   

Abstract

The Diabetes Technology Society surveillance protocol provides a seal of approval for a glucose meter if a sufficient number of a candidate glucose meter's results meet ISO 15197:2013 limits. The protocol provides clear details about how to conduct this study and analyze the data but has two flaws. There is no specification about the size of glucose meter errors that are outside of ISO limits. A meter that has a result in the E zone of a glucose meter error grid could receive the DTS seal of approval. In addition, the protocol uses the ISO standard, which could be considered a "state of the art" standard instead of an error grid, which is a clinical standard. Remedies for these problems are to replace the ISO standard with an error grid and to include requirements for errors found in C or higher zones of an error grid.

Entities:  

Keywords:  DTS Surveillance Protocol; ISO 15197:2013; error grid; performance specifications

Mesh:

Substances:

Year:  2017        PMID: 28376646      PMCID: PMC5951035          DOI: 10.1177/1932296817702926

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  6 in total

1.  Wrong thinking about glucose standards.

Authors:  Jan S Krouwer
Journal:  Clin Chem       Date:  2010-01-28       Impact factor: 8.327

2.  Defining analytical performance specifications: Consensus Statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine.

Authors:  Sverre Sandberg; Callum G Fraser; Andrea Rita Horvath; Rob Jansen; Graham Jones; Wytze Oosterhuis; Per Hyltoft Petersen; Heinz Schimmel; Ken Sikaris; Mauro Panteghini
Journal:  Clin Chem Lab Med       Date:  2015-05       Impact factor: 3.694

3.  The surveillance error grid.

Authors:  David C Klonoff; Courtney Lias; Robert Vigersky; William Clarke; Joan Lee Parkes; David B Sacks; M Sue Kirkman; Boris Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2014-06-13

Review 4.  Performance of Cleared Blood Glucose Monitors.

Authors:  David C Klonoff; Priya Prahalad
Journal:  J Diabetes Sci Technol       Date:  2015-06-08

Review 5.  A review of standards and statistics used to describe blood glucose monitor performance.

Authors:  Jan S Krouwer; George S Cembrowski
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

6.  Development of the Diabetes Technology Society Blood Glucose Monitor System Surveillance Protocol.

Authors:  David C Klonoff; Courtney Lias; Stayce Beck; Joan Lee Parkes; Boris Kovatchev; Robert A Vigersky; Guillermo Arreaza-Rubin; Robert D Burk; Aaron Kowalski; Randie Little; James Nichols; Matt Petersen; Kelly Rawlings; David B Sacks; Eric Sampson; Steve Scott; Jane Jeffrie Seley; Robbert Slingerland; Hubert W Vesper
Journal:  J Diabetes Sci Technol       Date:  2016-05-03
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.