| Literature DB >> 28367914 |
Tatsuya Tasaki1, Tomoko Ohshima2,3, Emi Usui1, Satoshi Ikawa4, Katsuhisa Kitano2,3, Nobuko Maeda2, Yasuko Momoi1.
Abstract
Non-mechanical procedures for removing caries-infected dentin are warranted in dentistry. We previously demonstrated the marked sterilization effect for direct irradiation of low-temperature plasma using dentin model infected with Streptococcus mutans. However it requires 180 s of intraoral plasma irradiation to eliminate bacteria. We alternatively investigated whether plasma-treated water (PTW), i.e., pure water exposed to plasma in an atmosphere, has a same bactericidal activity with the plasma irradiation. In the infected dentin model, the viable S. mutans counts recovered by bur at depth of 0.8-2.4 mm from the cavity floor were 104-106 CFU/round bur. After PTW application for only 10 s, the count was significantly decreased to below the detection limit (2.5 CFU/round bur) or 3.0±5.0 CFU/round bur. Since the bactericidal activity of PTW is rapidly deactivated at body temperature (37°C), PTW is likely to be biocompatible and holds significant potential for non-mechanical procedures for removing caries-infected dentin.Entities:
Keywords: Dentin; Plasma; Sterilization; Streptococcus mutans; Water
Mesh:
Substances:
Year: 2017 PMID: 28367914 DOI: 10.4012/dmj.2016-358
Source DB: PubMed Journal: Dent Mater J ISSN: 0287-4547 Impact factor: 2.102