A randomized comparative trial of OTSC and Padlock for upper GI hemostasis in a standardized experimental setting.

Ruediger L Prosst1, Thomas Kratt2.   

Abstract

INTRODUCTION: Upper gastrointestinal bleeding (UGIB) is the key emergency situation in clinical endoscopy and is traditionally treated with injection, thermal or through the scope clipping therapy. Mortality rates are in the range of 8-10% and demand new treatment approaches. The Over-The-Scope Clip (OTSC®) has been described as a very effective hemostatic device in UGIB. We compared OTSC with the Padlock™ device in an established pre-clinical setting.
MATERIAL AND METHODS: Our test-bed consisted of the biohybrid EASIE model using soft silicone tubes, tunneled into the gastric wall and surfacing at a mucosa defect, representing the bleeding site. After successful deployment of the OTSC and Padlock devices on the spurting ulcer bleed (Forrest Ia) the vessel tubes were pressurized with a manometer to 120 mmHg. Tight closure at this pressure was defined as successful hemostasis (primary endpoint). N = 11 procedures were done with each device. Statistical testing was done using Fisher's exact test. Sample size was adjusted to an assumed α-error of 5% (two-sided test) and a power of 80%.
RESULTS: Technically correct placement of the respective hemostatic device was achieved in all procedures. A statistically significant difference was found in the primary endpoint. In OTSC the success proportion was 100%; 11/11 (95% KI 74.1% to 100%); in Padlock it was 0%; 0/11 (95% KI 0%-25.8%). This means that no bleeding was stopped by Padlock. The mean value of perfusion pressure resistance was 300 mmHg (cut-off) for OTSC and 9.2 ± 8.4 mmHg for Padlock. DISCUSSION: Our data on hemostatic function of OTSC coincide with the clinical literature and earlier pre-clinical studies in the EASIE model, which is widely accepted as a realistic and effective simulation system for clinical conditions. The inability of Padlock to stop hemorrhage may be due to design differences and, thus, its limitation in providing tight sealing of the clipped tissue.
CONCLUSION: Different types of endoscope-tip mounted clips have different performances. OTSC consistently stops simulated spurting bleeding, Padlock fails to do so. These differences are statistically significant.

Entities:  

Keywords:  GI; OTSC; Padlock; bleeding; endoscopy

Mesh:

Year:  2017        PMID: 28350273     DOI: 10.1080/13645706.2017.1291523

Source DB:  PubMed          Journal:  Minim Invasive Ther Allied Technol        ISSN: 1364-5706            Impact factor:   2.442


  3 in total

1.  Reduced mortality for over-the-scope clips (OTSC) versus surgery for refractory peptic ulcer bleeding: a retrospective study.

Authors:  Armin Kuellmer; Tobias Mangold; Dominik Bettinger; Moritz Schiemer; Julius Mueller; Andreas Wannhoff; Karel Caca; Edris Wedi; Tobias Kleemann; Robert Thimme; Arthur Schmidt
Journal:  Surg Endosc       Date:  2022-10-17       Impact factor: 3.453

2.  Endoscopic Management with a Novel Over-The-Scope Padlock Clip System.

Authors:  Mahesh Kumar Goenka; Gajanan Ashokrao Rodge; Indrajeet Kumar Tiwary
Journal:  Clin Endosc       Date:  2019-11-26

3.  Over-the-scope clip versus transcatheter arterial embolization for refractory peptic ulcer bleeding-A propensity score matched analysis.

Authors:  Armin Kuellmer; Tobias Mangold; Dominik Bettinger; Lars Maruschke; Andreas Wannhoff; Karel Caca; Edris Wedi; Ali Seif Amir Hosseini; Tobias Kleemann; Thomas Schulz; Carlo Jung; Robert Thimme; Arthur Schmidt
Journal:  United European Gastroenterol J       Date:  2021-08-25       Impact factor: 4.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.