Literature DB >> 28346291

Round and Oval Window Anatomic Variability: Its Implication for the Vibroplasty Technique.

Marta Mancheño1, Miguel Aristegui, Jose Ramon Sañudo.   

Abstract

OBJECTIVE: The objective of this study is to evaluate the anatomical variability of round and oval window regions and its relationship with their closest structures, to determine its implication on the fitting and stabilization of the middle ear implant Vibrant Soundbridge.
METHODS: Variations of the anatomy of round and oval window regions were assessed in a total of 85 human dissected temporal bones. Afterward, we evaluated the adaptation and subsequent stabilization of the floating mass transducer (FMT) of the Vibrant Soundbridge in 67 cases in round window (RW) and in 22 cases in oval window (OW), and the influence that the variability of the different anatomical features examined had on this stabilization. We also assessed access and surgeon's view of the RW niche through the facial recess approach.
RESULTS: Stabilization of the FMT in the RW was achieved in 53 (79%) of the 67 cases; we found that the less favorable anatomical conditions for stabilization were: membrane smaller than 1.5 mm, presence of a high jugular bulb and a narrow or very narrow RW niche. Frequently, two or more of these conditions happened simultaneously. In seven cases (22%) access to the RW through facial recess approach did not allow positioning the FMT in place. OW stabilization succeeded in 18 (82%) of the 22 cases.
CONCLUSION: Round and oval window vibroplasty are difficult surgical techniques. To place the FMT directly on the OW may be easier as we do not have to drill the niche. In both regions there are some anatomical conditions that hinder fitting the FMT and even make it impossible. Once fitted, the main problem is to achieve good stabilization of the device.

Entities:  

Mesh:

Year:  2017        PMID: 28346291     DOI: 10.1097/MAO.0000000000001385

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  6 in total

1.  Anatomic features of the fetal round and oval windows, and their relations with the tympanic nerve.

Authors:  Orhan Beger; Yusuf Vayisoğlu; Onurhan Güven; Saliha Seda Adanır; Pourya Taghipour; Salim Çakır; Oykut Dağtekin; Derya Ümit Talas
Journal:  Surg Radiol Anat       Date:  2021-01-12       Impact factor: 1.246

2.  Anatomic variability of oval window as pertaining to stapes surgery.

Authors:  Anjali Singal; Daisy Sahni; Tulika Gupta; Anjali Aggarwal; Ashok Kumar Gupta
Journal:  Surg Radiol Anat       Date:  2019-09-23       Impact factor: 1.246

3.  The Hannover Coupler V2: Audiological outcomes of a round window coupler for the Vibrant Soundbridge.

Authors:  Nicole Knölke; Dawid Murawski; Nina Wardenga; Susan Busch; Hannes Maier; Thomas Lenarz
Journal:  Laryngoscope Investig Otolaryngol       Date:  2022-05-28

4.  Oval Window Size and Shape: a Micro-CT Anatomical Study With Considerations for Stapes Surgery.

Authors:  Matthew J Zdilla; Janusz Skrzat; Magdalena Kozerska; Bartosz Leszczyński; Jacek Tarasiuk; Sebastian Wroński
Journal:  Otol Neurotol       Date:  2018-06       Impact factor: 2.311

Review 5.  Round Window Stimulation of the Cochlea.

Authors:  Herman A Jenkins; Nathaniel Greene; Daniel J Tollin
Journal:  Front Neurol       Date:  2021-12-14       Impact factor: 4.003

6.  Redesign of the Hannover Coupler: Optimized Vibration Transfer from Floating Mass Transducer to Round Window.

Authors:  Mathias Müller; Rolf Salcher; Nils Prenzler; Thomas Lenarz; Hannes Maier
Journal:  Biomed Res Int       Date:  2018-05-14       Impact factor: 3.411

  6 in total