Can a commercially available EPID dosimetry system detect small daily patient setup errors for cranial IMRT/SRS?

Emmelyn S Hsieh1, Katherine S Hansen2, Michael S Kent2, Sanjeev Saini3, Sonja Dieterich4.   

Abstract

PURPOSE: The purpose of this study was to determine if the Sun Nuclear PerFRACTION electronic portal imager device dosimetry software would be able to detect setup errors in a clinical setting and would be able to correctly identify the direction in which the setup error was introduced. METHODS AND MATERIALS: A 7-field intensity modulated radiation therapy (IMRT) treatment plan for a centrally located tumor was developed for 1 phantom and 5 canine cadaver heads. Systematic setup errors were introduced by manually moving the treatment couch by 1, 3, and 5 mm in each translational direction to assess stereotactic radiation surgery (SRS), IMRT, and 3-dimensional (3D) treatment tolerances after the initial alignment was performed. An angular setup error of 5° yaw was also assessed. The delivered treatment fluence was automatically imported in the PerFRACTION software and compared with the baseline fluence.
RESULTS: In the canine phantom, a 5-mm shift was undetected by gamma analysis, and up to a 2-cm shift had to be introduced for the gamma pass rate of 3%/3 mm to fall below a 95% pass rate criterion. The same 5-mm shift using 3% difference caused the pass rates for 2 fields to drop below the 95% tolerance. For each respective translational shift, the affected beam angles were consistent across the cadaver heads and correlated with the direction of translational shift. The best field pass rate, worst field pass rate, and average pass rate across all 7 fields was analyzed to develop clinical guidance on parameter settings for SRS, IMRT, and 3D tolerances.
CONCLUSIONS: PerFRACTION 2-dimensional mode successfully detected setup errors outside the systematic error tolerance for SRS, IMRT, and 3D when an appropriate analysis metric and pass/fail criteria was implemented. Our data confirm that percent difference may be more sensitive in detecting plan failure than gamma analysis.
Copyright © 2017 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 28336480     DOI: 10.1016/j.prro.2016.12.005

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  7 in total

Review 1.  Emerging role of MRI in radiation therapy.

Authors:  Hersh Chandarana; Hesheng Wang; R H N Tijssen; Indra J Das
Journal:  J Magn Reson Imaging       Date:  2018-09-08       Impact factor: 4.813

2.  Evaluation of two-dimensional electronic portal imaging device using integrated images during volumetric modulated arc therapy for prostate cancer.

Authors:  Shoki Inui; Yoshihiro Ueda; Shunsuke Ono; Shingo Ohira; Masaru Isono; Yuya Nitta; Hikari Ueda; Masayoshi Miyazaki; Masahiko Koizumi; Teruki Teshima
Journal:  Rep Pract Oncol Radiother       Date:  2021-04-14

3.  Improving accuracy for stereotactic body radiotherapy treatments of spinal metastases.

Authors:  James Rijken; Barry Jordan; Scott Crowe; Tanya Kairn; Jamie Trapp
Journal:  J Appl Clin Med Phys       Date:  2018-06-26       Impact factor: 2.102

Review 4.  Imaging in Non-neurologic Oncologic Treatment Planning of the Head and Neck.

Authors:  Katherine S Hansen; Michael S Kent
Journal:  Front Vet Sci       Date:  2019-03-28

5.  Using Cherenkov imaging to monitor the match line between photon and electron radiation therapy fields on biological tissue phantoms.

Authors:  Yi Li; Hongjun Liu; Nan Huang; Zhaolu Wang; Chunmin Zhang
Journal:  J Biomed Opt       Date:  2020-12       Impact factor: 3.170

6.  Analysis of EPID Transmission Fluence Maps Using Machine Learning Models and CNN for Identifying Position Errors in the Treatment of GO Patients.

Authors:  Guyu Dai; Xiangbin Zhang; Wenjie Liu; Zhibin Li; Guangyu Wang; Yaxin Liu; Qing Xiao; Lian Duan; Jing Li; Xinyu Song; Guangjun Li; Sen Bai
Journal:  Front Oncol       Date:  2021-09-14       Impact factor: 6.244

7.  EPID-based daily verification of reproducibility of patients' irradiation with IMRT plans.

Authors:  Marta Kruszyna-Mochalska
Journal:  Rep Pract Oncol Radiother       Date:  2018-08-08
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.