Carotid artery bursting pressure and seal time after multiple uses of a vessel sealing device.

Jocelin S Blake1, Brian J Trumpatori2, Kyle G Mathews1, Emily H Griffith3.   

Abstract

OBJECTIVE: To determine any loss of function of a single-use bipolar vessel sealing device after repeated sterilization and use on canine carotid arteries. STUDY
DESIGN: Ex vivo; randomized block design (dog). SAMPLE POPULATION: Carotid artery segments (n = 90) from 15 canine cadavers.
METHODS: Six 10-mm LigaSure Atlas 20 cm devices (Covidien Inc, Minneapolis, Minnesota) were used to seal canine carotid artery segments in one cadaver, then resterilized using ethylene oxide and reused in a second cadaver, and so on, for a total of 15 uses/15 cadavers per LigaSure. Sealing times and bursting pressures of each carotid artery segment after each use were recorded. Charring and sticking at each seal location was qualitatively assessed.
RESULTS: Mean (SD) sealing time for all segments was 5.3 (1.4) seconds, ranging from 2.7 to 9.5 seconds. There was no significant change in sealing time for successive uses (P = .117). The mean (SD) bursting pressure for all segments was 1041.3 (316.7) mm Hg, ranging from 355 to 1555 mm Hg. There was no change in bursting pressure across uses (P = .57). Moderate sticking and charring occurred at each time point but was not subjectively different throughout the study.
CONCLUSION: The LigaSure Atlas vessel sealing device can be used and resterilized with ethylene oxide up to 15 times with no change in sealing time or bursting pressure of the seal.
© 2017 The American College of Veterinary Surgeons.

Entities:  

Mesh:

Year:  2017        PMID: 28334448     DOI: 10.1111/vsu.12648

Source DB:  PubMed          Journal:  Vet Surg        ISSN: 0161-3499            Impact factor:   1.495


  2 in total

1.  Use of a bipolar sealing device to seal partial cystectomy with and without augmentation with a single-layer simple continuous suture pattern in an ex vivo canine model.

Authors:  Valery F Scharf; Milan Milovancev; Katy L Townsend; Jennifer A Luff
Journal:  Vet Surg       Date:  2020-05-09       Impact factor: 1.495

2.  Effect of multiple use and sterilization on sealing performance of bipolar vessel sealing devices.

Authors:  Silvia Gardeweg; Barbara Bockstahler; Gilles Duprè
Journal:  PLoS One       Date:  2019-08-20       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.