[Intravenous digital subtraction angiography of the lower extremities combined with step translation (2-field DSA)].

P Fritschy1, F Terrier.   

Abstract

We evaluated a bolus-chase technique designed to reduce the volume of contrast material and to shorten examination time during intravenous digital subtraction angiography (IV-DSA) of the lower limbs. With this technique, after a single injection of contrast material, two contiguous fields can be sequentially imaged (two-field DSA) using a step translated x-ray tube. 67 patients with peripheral vascular disease were studied. All examinations provided diagnostic information, so that in no instance additional intra-arterial injections were required. With the two-field DSA only 3.49 injections were needed in the average for five fields examined in each patient, resulting in a 30% reduction of contrast agent used.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 2832892     DOI: 10.1055/s-2008-1048193

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  1 in total

1.  Adaptive Bolus Chasing Computed Tomography Angiography: Control Scheme and Experimental Results.

Authors:  Zhijun Cai; Colbin Erdahl; Kai Zeng; Tom Potts; Melhem Sharafuddin; Osama Saba; Ge Wang; Er-Wei Bai
Journal:  Biomed Signal Process Control       Date:  2008-10       Impact factor: 3.880