| Literature DB >> 28321836 |
D Onwochei1, K El-Boghdadly1, R Oakley1, I Ahmad1.
Abstract
We present the case of unanticipated airway ignition during hard palate biopsy. Transnasal humidified rapid-insufflation ventilatory exchange (THRIVE) and monopolar diathermy were utilised for the procedure, during which an arc arose from the diathermy tip to a titanium implant, causing a brief ignition on the monopolar diathermy grip. This case highlights the need for maintained awareness of fire risk when using diathermy in the presence of THRIVE during airway surgery.Keywords: airway fire; airway ignition; diathermy; oral surgery
Mesh:
Substances:
Year: 2017 PMID: 28321836 DOI: 10.1111/anae.13873