An In Vitro Assessment of the Efficacy of Various IV Cannulas for the Rapid IV Fluid Administration.

Mineto Kamata1, Hina Walia, Mumin Hakim, Dmitry Tumin, Joseph D Tobias.   

Abstract

OBJECTIVES: The current study prospectively evaluates the administration of fluid through commonly used vascular cannulas of various length and diameter.
DESIGN: Observational, in vitro experiment.
SETTING: Not applicable.
SUBJECTS: No human subjects.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: Fluid (500 mL) was administered via gravity flow and with pressure assistance (pressure bag set at 300 mm Hg) through various vascular cannulas including peripheral IV catheters (22 gauge, 1 inch; 20 gauge, 1.16 and 1.88 inch; and 18 gauge, 1.16 and 1.88 inch), 3-Fr central line lumens (lengths 50, 80, and 120 mm), and a 4-Fr, 10 cm Micropuncture catheter (Cook Medical, Bloomington, IN). During gravity flow, drain time decreased by approximately 50% when moving from a 22 to 20 to 18 gauge cannula and increased by approximately 20% as the catheter length increased from 50 to 80 to 120 cm. Flow rates were highest with the Micropuncture catheter, which achieved a drain time of 2.7 minutes for the 500 mL bag when infused with pressure assistance.
CONCLUSIONS: In general, the delivery of crystalloid solutions through commonly used vascular cannulas was fastest with larger diameter and shorter length cannulas. Pressure-assisted flow was able to partially compensate for the increased resistance with smaller and longer catheters. The unique design of the tip of the Micropuncture catheter compensated for the increased length and allowed for rapid fluid delivery. This design may compensate for the increased length when longer catheters are needed for ultrasound-guided placement.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28319488     DOI: 10.1097/PCC.0000000000001151

Source DB:  PubMed          Journal:  Pediatr Crit Care Med        ISSN: 1529-7535            Impact factor:   3.624


  3 in total

1.  A novel approach to obtaining large-bore access in pediatrics.

Authors:  Mathew B Kiberd; Tristan Dumbarton
Journal:  Can J Anaesth       Date:  2022-09-30       Impact factor: 6.713

2.  [Application of ultrasound-guided central venous catheterization at various sites in infants with shock].

Authors:  Zi-Feng Tan; Ke-Ze Ma; Zhi-Jun Lai
Journal:  Zhongguo Dang Dai Er Ke Za Zhi       Date:  2022-05-15

3.  Rapid fluid administration: an evaluation of two techniques.

Authors:  Holly C Gillis; Hina Walia; Dmitry Tumin; Tarun Bhalla; Joseph D Tobias
Journal:  Med Devices (Auckl)       Date:  2018-09-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.