| Literature DB >> 28319139 |
Saeam Shin1,2, Sung Ran Cho3, Sinyoung Kim1, Jong Rak Choi1, Kyung-A Lee1.
Abstract
Optimal timing of apheresis initiation is important for maximizing the hematopoietic stem cell (HSC) yield. This study aimed to identify useful parameters from automatic hematology analyzers for predicting the peripheral blood CD34+ cell count after mobilization. We prospectively enrolled 53 healthy donors and 72 patients, and evaluated 43 cell morphology parameters from Unicel DxH800 (Beckman Coulter, USA) and Advia 2120i (Siemens Healthcare Diagnostics, USA). The correlation of each parameter with the CD34+ cell count in pre-apheresis blood samples was analyzed. The delta neutrophil index (DNI) from Advia 2120i, standard deviation of volume of neutrophils and monocytes (SD-V-NE and SD-V-MO), standard deviation of conductivity of neutrophils and monocytes (SD-C-NE and SD-C-MO), mean conductivity of neutrophils and monocytes (MN-C-NE and MN-C-MO), and standard deviation of axial light loss of neutrophils and monocytes (SD-AL2-NE and SD-AL2-MO) from DxH800 showed significant correlations with the CD34+ cell count. SD-V-NE, SD-C-NE, and SD-C-MO showed good or fair area under the curve values for the prediction of the CD34+ cell count. SD-V-NE, SD-C-NE, and SD-C-MO from DxH800 will provide rapid, useful information for the initiation of apheresis after mobilization.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28319139 PMCID: PMC5358893 DOI: 10.1371/journal.pone.0174286
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Scatter plots and Pearson’s correlation analyses between (A) PB CD34+ cells/μL and CD34+ cells/μL in a product, and (B) PB CD34+ cells/μL and total CD34+ cells/kg donor body weight (×106).
PB, peripheral blood.
Receiver operating characteristics analysis using cell morphology parameters for prediction of the CD34+ cell count.
| AUC for target CD34+ cells/μL in PB | ||||||
|---|---|---|---|---|---|---|
| Parameters | P-value | < 15, ≥ 15 (n=20, 105) | < 20, ≥ 20 (n=25, 100) | < 25, ≥ 25 (n=31, 94) | < 30, ≥ 30 (n=39, 86) | |
| DNI | 0.416 | < 0.001 | 0.598 | 0.587 | 0.617 | 0.633 |
| SD-V-NE | 0.313 | < 0.001 | 0.697 | 0.716 | 0.728 | 0.695 |
| SD-V-MO | 0.289 | 0.001 | 0.691 | 0.624 | 0.631 | 0.573 |
| SD-C-NE | 0.285 | 0.001 | 0.845 | 0.778 | 0.769 | 0.704 |
| MN-C-NE | -0.275 | 0.002 | 0.588 | 0.563 | 0.553 | 0.540 |
| SD-C-MO | 0.256 | 0.004 | 0.761 | 0.699 | 0.705 | 0.648 |
| SD-AL2-MO | 0.246 | 0.006 | 0.606 | 0.574 | 0.603 | 0.573 |
| SD-AL2-NE | 0.226 | 0.01 | 0.535 | 0.507 | 0.530 | 0.506 |
| MN-C-MO | -0.197 | 0.03 | 0.625 | 0.606 | 0.567 | 0.515 |
AUC, area under the receiver operating characteristics curve; PB, peripheral blood; DNI, delta neutrophil index; SD, standard deviation; V, volume; NE, neutrophil; MO, monocyte; C, conductivity; MN, mean; AL2, axial light loss.
a The remaining 34 parameters with a P-value of Pearson’s correlation test ≥ 0.05 were omitted.
b Pearson's correlation coefficient between parameters from hematologic analyzers and peripheral blood CD34+ cell count.