| Literature DB >> 28299804 |
Huixiang Liu1, Meitong Zang1, Aijuan Yang1, Jianbo Ji1, Jie Xing1.
Abstract
A sensitive and efficient liquid chromatography tandem mass spectrometry method was developed and validated for the simultaneous determination of piperaquine (PQ) and its N-oxidated metabolite (PQ-M) in plasma. A simple protein precipitation procedure was used for sample preparation. Adequate chromatographic retention was achieved on a C18 column under gradient elution with acetonitrile and 2 mm aqueous ammonium acetate containing 0.15% formic acid and 0.05% trifluoroacetic acid. A triple-quadrupole mass spectrometer equipped with an electrospray source was set up in the positive ion mode and multiple reaction monitoring mode. The method was linear in the range of 2.0-400.0 ng/mL for PQ and 1.0-50.0 ng/mL for PQ-M with suitable accuracy, precision and extraction recovery. The lower limits of detection (LLOD) were established at 0.4 and 0.2 ng/mL for PQ and PQ-M, respectively, using 40 μL of plasma sample. The matrix effect was negligible under the current conditions. No effect was found for co-administrated artemisinin drugs or hemolysis on the quantification of PQ and PQ-M. Stability testing showed that two analytes remained stable under all relevant analytical conditions. The validated method was successfully applied to a pharmacokinetic study performed in rats after a single oral administration of PQ (60 mg/kg).Entities:
Keywords: LC-ESI-MS/MS; metabolite; pharmacokinetics; piperaquine; rat plasma
Mesh:
Substances:
Year: 2017 PMID: 28299804 DOI: 10.1002/bmc.3974
Source DB: PubMed Journal: Biomed Chromatogr ISSN: 0269-3879 Impact factor: 1.902