Literature DB >> 28296065

Direct percutaneous transaxillary implantation of a novel self-expandable transcatheter heart valve for aortic stenosis.

Florian Deuschl1, Niklas Schofer1, Moritz Seiffert1, Samer Hakmi2, Isamu Mizote1, Andreas Schaefer2, Johannes Schirmer2, Hermann Reichenspurner2, Stefan Blankenberg1, Lenard Conradi2, Ulrich Schäfer1.   

Abstract

OBJECTIVES: The aim of this study was to evaluate safety, feasibility, and efficacy of transaxillary TAVI using a novel self-expandable transcatheter heart valve (THV) via a direct percutaneous technique.
METHODS: The analysis comprises 12 consecutive patients with severe aortic stenosis treated with the ACURATE Neo™ THV (Symetis) by direct percutaneous transaxillary access.
RESULTS: Patients presented with a mean age of 79.9 ± 6.2 years and high calculated surgical risk (logEuroSCORE I: 19.55 ± 15.00%; STS PROM score: 5.42 ± 3.25%). Transaxillary TAVI was performed in 11/12 via the left and in 1/12 via the right-sided axillary artery. Device success was achieved in all patients (12/12). Mortality rate at 30 days was 0% (0/12). No major and one minor access site complication (1/12) occurred. No major or life-threatening access site-related bleeding resulted. None of the patients suffered from postprocedural stroke/transient ischemic attack. There was no need for a permanent pacemaker. Mean effective orifice area post implantation was 2.1 ± 0.29 cm2 and mean aortic gradient 6.2 ± 1.5 mm Hg. In none of the patients, greater than moderate aortic regurgitation or paravalvular leakage was detected at discharge (0/12).
CONCLUSIONS: Our study suggests that the implantation of the second-generation self-expandable ACURATE Neo™ via a transaxillary access is technically feasible and safe and yields excellent short-term clinical results.
© 2017 Wiley Periodicals, Inc. © 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  TAVI; THV; alternative access; transaxillary access

Mesh:

Year:  2017        PMID: 28296065     DOI: 10.1002/ccd.26986

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  2 in total

1.  Percu-Ax aortic valve implantation with a double arm approach: a case report.

Authors:  Carlo Trani; Cristina Aurigemma; Enrico Romagnoli; Francesco Burzotta
Journal:  Eur Heart J Case Rep       Date:  2020-10-16

2.  A meta-analysis comparing transaxillary and transfemoral transcatheter aortic valve replacement.

Authors:  Yong Zhan; Siavash Saadat; Avneet Soin; Masashi Kawabori; Frederick Y Chen
Journal:  J Thorac Dis       Date:  2019-12       Impact factor: 2.895

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.