Literature DB >> 28295385

Technical Note: The design and function of a horizontal patient rotation system for the purposes of fixed-beam cancer radiotherapy.

Ilana Feain1, Lloyd Coleman2, Hue Wallis2, Richard Sokolov2, Ricky O'Brien1, Paul Keall1.   

Abstract

PURPOSE: Cancer radiation therapy treatment is performed by delivering a 3D dose distribution to the tumor via the relative rotation between beam and patient. While most modern machines rotate the radiation beam around a still patient, the treatment can also be delivered by rotating the patient relative to a fixed beam. Fixed-beam, patient rotation radiotherapy machines show promise for reducing the size, surface area footprint, and shielding requirements compared with rotating gantry machines. In this Technical Note, we describe the development of a bespoke horizontal patient rotation system for the purposes of a fixed-beam cancer radiotherapy architecture.
METHODS: A horizontal Patient Rotation System was designed in accordance with the appropriate standards pertaining to performance and safety of medical electrical equipment and medical linear accelerators (ISO 9001, IEC 60601-1, IEC 60601-2-1, ISO 14971, ISO 13485, 21CFR820, IEC 62304, Machinery Directive 98/37/EC). The principal criteria for the design were safety, patient comfort, real-time control and the ability to be integrated with other radiation therapy componentry (including a linear accelerator and kV imaging systems).
RESULTS: A first of its kind device for securing, immobilizing, translating, and rotating patients has been designed and built and tested against 161 different design, safety, and usability specifications. The device has real-time control for all critical applications.
CONCLUSIONS: We designed and built a bespoke device which can translate and rotate patients 360° around a horizontal axis. The device meets all design and safety criteria with early usability tests indicating a high degree of comfort and utility. The system has been installed in a clinical bunker, integrated with a fixed-beam linear accelerator and is currently being commissioned for the purposes of cancer radiotherapy treatment.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  fixed-beam cancer radiotherapy; patient rotation; radiotherapy

Mesh:

Year:  2017        PMID: 28295385     DOI: 10.1002/mp.12219

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Technical Note: Density correction to improve CT number mapping in thoracic deformable image registration.

Authors:  Jinzhong Yang; Yongbin Zhang; Zijian Zhang; Lifei Zhang; Peter Balter; Laurence Court
Journal:  Med Phys       Date:  2019-04-01       Impact factor: 4.071

2.  Please Place Your Seat in the Full Upright Position: A Technical Framework for Landing Upright Radiation Therapy in the 21st Century.

Authors:  Sarah Hegarty; Nicholas Hardcastle; James Korte; Tomas Kron; Sarah Everitt; Sulman Rahim; Fiona Hegi-Johnson; Rick Franich
Journal:  Front Oncol       Date:  2022-03-03       Impact factor: 6.244

  2 in total