An In Vitro Evaluation of the Loosening of Different Interchangeable Abutments in Internal-Connection-Type Implants.

Ji-Man Park, Chang-Hyun Baek, Seong-Joo Heo, Seong-Kyun Kim, Jai-Young Koak, Shin-Koo Kim, Urs C Belser.   

Abstract

PURPOSE: The aim of this study was to compare the loosening of interchangeable one-piece abutments connected to internal-connection-type implants after cyclic loading.
MATERIALS AND METHODS: Four implant abutment groups (n = 7 in each group) with Straumann tissue-level implants were assessed: Straumann solid abutment (group S), Southern Implants solid abutment (group SI), Implant Direct straight abutment (group ID), and Blue Sky Bio regular platform abutment (group BSB). The implant was firmly held in a special jig to ensure fixation. Abutment screws were tightened to manufacturers' recommended torque with a digital torque gauge. The hemispherical loading members were fabricated for the load cell of a universal testing machine to evenly distribute the force on the specimens and to fulfill the ISO 14801:2007 standard. A cyclic loading of 25 N at 30 degrees to the implant's long axis was applied for a duty of a half million cycles. Tightening torques were measured prior to the loading. Removal torques were measured after cyclic loading. The data were analyzed with one-way analysis of variance (ANOVA), and the significance level was set at P < .05.
RESULTS: The mean removal torques after cyclic loading were 34.0 ± 1.1 Ncm (group S), 25.0 ± 1.5 Ncm (group SI), 23.9 ± 2.1 Ncm (group ID), and 27.9 ± 1.3 Ncm (group BSB). Removal torques of each group were statistically different in the order of group S > group BSB > groups SI and ID (P < .05). The mean reduction rates were -2.9% ± 3.2% (group S), -21.9% ± 4.8% (group SI), -20.2% ± 7.2% (group ID), and -6.9% ± 4.3% (group BSB) after a half million cycles, respectively. Reduction rates of groups S and BSB were statistically lower than those of groups SI and ID (P < .01). The standard deviation of group S was lower than group BSB.
CONCLUSION: The removal torque of the original Straumann abutment was significantly higher than those of the copy abutments. The reduction rate of the groups S and BSB abutments was lower than those of the other copy abutments.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28291853     DOI: 10.11607/jomi.5295

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  5 in total

Review 1.  Properties of polyetheretheretherketone (PEEK) implant abutments: A systematic review.

Authors:  Romaisa Ghazal-Maghras; Jaime Vilaplana-Vivo; Fabio Camacho-Alonso; Yolanda Martínez-Beneyto
Journal:  J Clin Exp Dent       Date:  2022-04-01

2.  A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis.

Authors:  Rimantas Ožiūnas; Jurgina Sakalauskienė; Darius Jegelevičius; Gintaras Janužis
Journal:  J Adv Prosthodont       Date:  2022-04-27       Impact factor: 1.989

3.  Mechanical Outcomes, Microleakage, and Marginal Accuracy at the Implant-Abutment Interface of Original versus Nonoriginal Implant Abutments: A Systematic Review of In Vitro Studies.

Authors:  Marco Tallarico; Joseph Fiorellini; Yasushi Nakajima; Yuki Omori; Iida Takahisa; Luigi Canullo
Journal:  Biomed Res Int       Date:  2018-12-30       Impact factor: 3.411

4.  Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force.

Authors:  Yuseung Yi; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

5.  Comparing the Fracture Resistance and Modes of Failure in Different Types of CAD/CAM Zirconia Abutments with Internal Hexagonal Implants: An In Vitro Study.

Authors:  Yu-Tsen Chang; Yu-Ling Wu; Hung-Shyong Chen; Ming-Hsu Tsai; Chia-Chen Chang; Aaron Yu-Jen Wu
Journal:  Materials (Basel)       Date:  2022-04-04       Impact factor: 3.623

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.