Poor Acetabular Component Orientation Increases Revision Risk in Metal-on-Metal Hip Arthroplasty.

Tuomas J T Tauriainen1, Tuukka T Niinimäki1, Jaakko L Niinimäki2, Tomi O P Nousiainen1, Juhana I Leppilahti1.   

Abstract

BACKGROUND: The rate of and the reasons for the failure of metal-on-metal (MoM) bearings have recently been discussed in literature. The aim of this study was to evaluate the influence of acetabular cup inclination and version angles on revision risk in patients with MoM hip arthroplasty.
METHODS: We retrospectively reviewed 825 patients (976 hips) who underwent a MoM hip arthroplasty between 2000 and 2013. There were 474 men and 351 women, with a mean age of 58 (19-86) years. Acceptable cup orientation was considered to be inside the Lewinnek's safe zone.
RESULTS: The mean acetabular inclination angle was 48.9° (standard deviation, 8.1°; range, 16°-76°) and version angle 20.6° (standard deviation, 9.9°; range, -25 to 46°). The cup was found to be outside the Lewinnek's safe zone in 571 hips (58.5%). Acetabular cup revision surgery was performed in 157 hips (16.1%). The cup angles were outside Lewinnek's safe zone in 69.2% of the revised hips. The mean interobserver reliability and intraobserver repeatability of the measurements of cup inclination and version angles were excellent (intraclass correlation coefficients >0.90). The odds ratio for revision in hips outside vs inside the Lewinnek's safe zone was 1.82 (95% confidence interval, 1.26-2.62; P = .0014).
CONCLUSION: Our findings provide compelling evidence that a cup position outside the Lewinnek's safe zone is associated with increased revision risk in patients with MoM arthroplasty.
Copyright © 2017. Published by Elsevier Inc.

Entities:  

Keywords:  Lewinnek's safe zone; acetabular component orientation; metal-on-metal; revision; total hip arthroplasty

Mesh:

Year:  2017        PMID: 28291650     DOI: 10.1016/j.arth.2017.02.009

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

1.  Lewinnek Safe Zone References are Frequently Misquoted.

Authors:  Aonnicha Burapachaisri; Ameer Elbuluk; Edem Abotsi; Jim Pierrepont; Seth A Jerabek; Aaron J Buckland; Jonathan M Vigdorchik
Journal:  Arthroplast Today       Date:  2020-11-26

2.  Uncemented versus cemented arthroplasty after metal-on-metal total hip replacement in patients with femoral neck fractures: a retrospective study.

Authors:  Wenlu Liu; Huanyi Lin; Xianshang Zeng; Meiji Chen; Weiwei Tang; Ting Zhou; Weiguang Yu; Qilong Liu; Guixing Xu
Journal:  J Int Med Res       Date:  2021-05       Impact factor: 1.671

3.  Retention of metals in periprosthetic tissues of patients with metal-on-metal total hip arthroplasty is reflected in the synovial fluid to blood cobalt transfer ratio in the presence of a pseudotumour.

Authors:  Tomi Nousiainen; Sanna Palosaari; Sirpa Peräniemi; Arja Tervahauta; Jaakko Niinimäki; Juhana Leppilahti; Petri Lehenkari
Journal:  BMC Musculoskelet Disord       Date:  2020-09-12       Impact factor: 2.362