Literature DB >> 28285931

Flow and wall shear stress characterization after endovascular aneurysm repair and endovascular aneurysm sealing in an infrarenal aneurysm model.

Johannes T Boersen1, Erik Groot Jebbink2, Michel Versluis3, Cornelis H Slump4, David N Ku5, Jean-Paul P M de Vries6, Michel M P J Reijnen7.   

Abstract

BACKGROUND: Endovascular aneurysm repair (EVAR) with a modular endograft has become the preferred treatment for abdominal aortic aneurysms. A novel concept is endovascular aneurysm sealing (EVAS), consisting of dual endoframes surrounded by polymer-filled endobags. This dual-lumen configuration is different from a bifurcation with a tapered trajectory of the flow lumen into the two limbs and may induce unfavorable flow conditions. These include low and oscillatory wall shear stress (WSS), linked to atherosclerosis, and high shear rates that may result in thrombosis. An in vitro study was performed to assess the impact of EVAR and EVAS on flow patterns and WSS.
METHODS: Four abdominal aortic aneurysm phantoms were constructed, including three stented models, to study the influence of the flow divider on flow (Endurant [Medtronic, Minneapolis, Minn], AFX [Endologix, Irvine, Calif], and Nellix [Endologix]). Experimental models were tested under physiologic resting conditions, and flow was visualized with laser particle imaging velocimetry, quantified by shear rate, WSS, and oscillatory shear index (OSI) in the suprarenal aorta, renal artery (RA), and common iliac artery.
RESULTS: WSS and OSI were comparable for all models in the suprarenal aorta. The RA flow profile in the EVAR models was comparable to the control, but a region of lower WSS was observed on the caudal wall compared with the control. The EVAS model showed a stronger jet flow with a higher shear rate in some regions compared with the other models. Small regions of low WSS and high OSI were found near the distal end of all stents in the common iliac artery compared with the control. Maximum shear rates in each region of interest were well below the pathologic threshold for acute thrombosis.
CONCLUSIONS: The different stent designs do not influence suprarenal flow. Lower WSS is observed in the caudal wall of the RA after EVAR and a higher shear rate after EVAS. All stented models have a small region of low WSS and high OSI near the distal outflow of the stents.
Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28285931     DOI: 10.1016/j.jvs.2016.10.077

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  4 in total

1.  Experimental Analysis of Pressure and Flow Alterations During and After Insertion of a Multilayer Flow Modulator into an AAA Model with Incorporated Branch.

Authors:  Simon Tupin; Kei Takase; Makoto Ohta
Journal:  Cardiovasc Intervent Radiol       Date:  2021-04-27       Impact factor: 2.740

2.  Endovascular Aneurysm Sealing is Associated with Higher Medium-Term Survival than Traditional EVAR.

Authors:  Thomas F X O'Donnell; Jeffrey P Carpenter; John S Lane; Jose Trani; Sajjad Hussain; Christopher Healey; Mahmoud B Malas; Marc L Schermerhorn
Journal:  Ann Vasc Surg       Date:  2019-10-16       Impact factor: 1.466

3.  Editorial: Clinical and Hemodynamic Performance of Aortic Endografts.

Authors:  Efstratios Georgakarakos
Journal:  Front Surg       Date:  2018-06-22

4.  A novel roller pump for physiological flow.

Authors:  Albert Chong; Zhonghua Sun; Lennart van de Velde; Shirley Jansen; Michel Versluis; Michel M P J Reijnen; Erik Groot Jebbink
Journal:  Artif Organs       Date:  2020-03-25       Impact factor: 3.094

  4 in total