Literature DB >> 28282027

Feasibility of the backprojection method for reconstruction of low contrast perturbations in a conducting background in magnetic induction tomography.

A V Korjenevsky1, S A Sapetsky.   

Abstract

OBJECTIVE: Application of the backprojection image reconstruction algorithm in magnetic induction tomography for imaging of low contrast inclusions in conductive background is investigated. APPROACH: Both FDTD numerical simulation and experiments with the 16-channel measuring system and saline phantom are used. MAIN
RESULTS: It is demonstrated that spatial redistribution of the sensitivity of measurements related with background conductivity does not affect the projection lines and keeps possibility of backprojection image reconstruction, which is important in biomedical applications of MIT. SIGNIFICANCE: The applicability of backprojection in biomedical MIT has been proved. The reconstruction algorithm is explained in detail.

Mesh:

Year:  2017        PMID: 28282027     DOI: 10.1088/1361-6579/aa6605

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  1 in total

1.  Sector sensor array technique for high conductivity materials imaging in magnetic induction tomography.

Authors:  Jia Chen; Li Ke; Qiang Du; Wanni Zu; Xiaodi Ding
Journal:  Biomed Eng Online       Date:  2019-12-02       Impact factor: 2.819

  1 in total