Nematollah Heidarloo1, Hamid Reza Baghani2, Seyed Mahmoud Reza Aghamiri1, Seyed Rabi Mahdavi3, Mohammad Esmail Akbari4.
Abstract
Beam shaper is a newly designed device for beam shaping in IOERT. The aim of this study is evaluating the performance of this device for conformal IOERT and preparing it for clinical applications. Dosimetric characteristics for different combinations of energy/field size were determined by ionometric dosimetry, film dosimetry and Monte Carlo simulation. Desirable dosimetric characteristics of beam shaper make it a useful tool for conformal IOERT. Usability of Monte Carlo simulation in any clinical setup of beam shaper was demonstrated.Keywords: Beam shaper applicator; Electron dosimetry; IOERT; Monte carlo simulation
Mesh:
Year: 2017 PMID: 28260609 DOI: 10.1016/j.apradiso.2017.02.039
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513