Literature DB >> 28258831

Greater Trochanteric Reattachment Using the Third-Generation Cable Plate System in Revision Total Hip Arthroplasty.

In-Sung Kim1, Nachiket Pansey1, Rajeev K Kansay1, Je-Hyun Yoo2, Hwang-Yong Lee1, Jun-Dong Chang1.   

Abstract

BACKGROUND: The purpose of the study is to evaluate the radiologic and functional results of greater trochanteric reattachment using the third-generation cable plate system in revision total hip arthroplasty (THA).
METHODS: A total of 47 trochanteric fixations (27 men and 18 women; mean age of 60.2 years) using the third-generation cable plate system in revision THA were retrospectively evaluated. The mean follow-up was 80.4 months (range 27-148 months). The osteotomized greater trochanter was reattached using the Cable-Ready system (Zimmer, Warsaw, IN) and the Dall-Miles cable system (Stryker, Mahwah, NJ). The clinical results with Harris hip score, visual analog scale, and radiologic outcomes were evaluated.
RESULTS: The mean Harris hip score was improved from 55.7 (range 17-72) preoperatively to 90.8 (range 68-100; P = .001) postoperatively, and the mean pain score was improved from 6.6 (range 3-10) to 2.5 (range 0-6; P = .001), respectively. Nonunion was observed in 6 hips (12.7%). Migration of the osteotomized greater trochanteric fragment (>1 cm) was seen in 8 hips (17.0%). Cable breakage occurred in 13 cases (27.6%). Although 5 cable plate systems were removed, there was no need for reattachment of the greater trochanter in this study.
CONCLUSION: This study showed a relatively high incidence of radiologic failure after greater trochanteric reattachment using the cable plate system in revision THA, although reattachments were not required and clinical outcome was relatively satisfactory. Periodic and close observation for the early detection of failure is necessary.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  arthroplasty; cable plate; fixation device; greater trochanter; revision; trochanteric osteotomy

Mesh:

Year:  2017        PMID: 28258831     DOI: 10.1016/j.arth.2017.01.017

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  4 in total

1.  Greater trochanter fixed with a claw plate and cable system in complex primary and revision total hip arthroplasty: long-term follow-up.

Authors:  Jing Tang; Tiemure Wu; Hongyi Shao; Yixin Zhou
Journal:  Int Orthop       Date:  2022-08-23       Impact factor: 3.479

2.  Fixation of Trochanteric Fragments in Cementless Bipolar Hemiarthroplasty of Unstable Intertrochanteric Fracture: Cerclage Wiring.

Authors:  Young-Kyun Lee; Chan Ho Park; Kyung-Hoi Koo
Journal:  Hip Pelvis       Date:  2017-12-01

3.  The Relationship Between Arthroplasty Surgeons' Experience Level and Optimal Cable Tensioning in the Fixation of Extended Trochanteric Osteotomy.

Authors:  Kerem Başarır; Mahmut Kalem; Ercan Şahin; Emre Anıl Özbek; Mustafa Onur Karaca; İbrahim Küçükkarapınar; Ergin Tönük
Journal:  Geriatr Orthop Surg Rehabil       Date:  2021-12-13

Review 4.  Trochanteric osteotomy in revision total hip arthroplasty.

Authors:  Kavin Sundaram; Ahmed Siddiqi; Atul F Kamath; Carlos A Higuera-Rueda
Journal:  EFORT Open Rev       Date:  2020-09-10
  4 in total